    The HV software was optimized and tested in the DVCS area, and moved to R-HRS. Unfortunately, there are two problems.

    1. The upper crate, which has 15 cards, sometimes reports a few cards missing. Which ones are missing is somewhat random, but 0, 2, and 4 are most popular. This occurs with the old software, too. It does not occur with either old or new software on the bottom crate, which has 3 cards. I tried to power-cycle the crate, it did not help.

    2. The new software isn't working -- there is a different response from the DVCS crate. This is true for both top and bottom crate, using the Intel PC to talk. I suspect something about the Intel vs the DVCS PC. I'll investigate. The old software with long delays works, except for problem #1.