    keyword=DAQ issues

    I noticed a couple things
    
    1. On L-HRS, 43% of triggers taken at the moment are pulser T8.  This is
    because the 103.7 kHz clock is plugged in there, I think, and the
    prescale factor is too low for that.  For the max PS factor is 65535,
    you'll always have 1.6 Hz.
    
    2. On R-HRS, it seems ROC6 has a problem -- empty readout.  I didn't look
    into it further, probably a cable fell out or was removed.
    