    User name Kalyan

    Log entry time 10:35:11 on April 10, 2014

    Entry number 388674

    This entry is a followup to: 388673

    Followups:

    keyword=Bad 24 V power supply on RHRS HV crate?

    Attached is the screenshot of error messages from right arm HV crate on
    ahut2 (which has only 4 cards). The channels trip when HV is applied.
    This happens on all four cards. We disconnected the HV cables from the
    detector (VDC on slot 12) and tried to apply HV from the GUI - the HV
    still trips. The error messages says BAD POWER SUPPLY 24V and BAD EEPROM. 
    
    I restarted the HV server from ahut2, just to make sure if this is not a
    software issue. We still see HV tripping. It looks like the problem is
    related to crate power supply?
    
    


    A copy of this log entry has been emailed to: yangw,bogdanw,rom