User name mongi
Log entry time 20:15:47 on November14,2014
Entry number 404010
keyword=Target Loop1 temperature after heater does not look correct
The temperature difference of the target after the heater does not look correct for loop 1. After heater, temperatur is lower than it is for the liquid ibefore the heater. It looks like the temperature display is swapped or the gas flow is not as suggested by the diagram.
With J.P Chen, we reduced the power of the heater to zero and we still observed a 0.16K difference. J.P thinks this might be an artifact of some offset in the temperature sensors.
FIGURE 1