• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name bogdanw

    Log entry time 19:31:07 on September 27,1998

    Entry number 1664

    keyword=HA VDC

    On HA VDC had a problem of hot wires ~ numbers 290-300 on planes 3 and 4.
    This time we ( Segal, Prout and me ) made the following:
    1. cleanup all connections of the amplifiers to the wires of the TOP chamber ,
    2. replaced two amplifiers U5 and V5 which were hot wires,
    3. fixed weak connection on the power supply and threshold lines,
    4. put threshold 3.00 V on both chambers (TOP and BOT).
    5. checkout on cosmic ~ 1/2 hour run #1294- looks good.