User name J. Gomez
Log entry time 10:33:15 on October 27,1998
Entry number 3824
This entry is a followup to: 3782
keyword=SPECTROMETER ALIGNMENT
Regarding the medm alignment screen for both electron/hadron spectrometers:
There are two fields (yellow color) labeled "Flr Mrk (deg)" and "Vernier (mm)"
that need to be filled before the code can calculate the spectrometer angle and
pointing. Obtain these values from the TV-monitors looking at the floor marks.
Please check on the TV-monitor screen the sign (+/-) of the vernier reading
(it will make a large difference).
Since the code needs the above mentioned fields before a calculation can be
performed, the screen is blanked out and ERROR# 2 is displayed. After a reboot
of the magnet IOC (either electron or hadron) these fields need to be re-entered.
Note that if this fields are not entered, the summary files written to HALOG will
also have a "0" for the corresponding spectrometer angle. Also, if the spectrometers
are moved, the new values of "Flr Mrk" and "Vernier" need to be entered
manually (THEY CAN NOT BE READ AUTOMATICALLY).