User name Eugene
Log entry time 23:18:06 on December 15,1998
Entry number 8713
Followups:
keyword=Moller quad settings
The Moller quads were set by MCC to :
MQM1H02 18050 100A
MQO1H03 -12630 75A
MQO1H03A 26840 160A
which is a unified setting for both Moller and regular running.
The power supplies for 1H02 were found turned off and 1H03A was at about
130A instead of 160A. It looks like this setting ran across some limitation,
though the limit for a couple of power supplies, working in parallel should be about
240A, well above 160A. EPICS somehow managed to read out 18A from
the turned off power supplies!
We decided to run this Moller measurement at a different
setting:
MQM1H02 8120 43A
MQO1H03 3610 20A
MQO1H03A 15810 93A
which are equally good for Moller and , accordingly to Jay Benesch ,
acceptable for regular running.
One should contact Jay and H.Areti tomorrow.