    Checked the timing on H-arm Scintillator S1 and S2
    paddle 2 input to the MLU relative to the strobe for
    MLU. This was checked on a scope in the counting
    room. It was the same timing as for paddle 4 checked
    earlier in the week. The timing looks adequate. This
    does not explain, however, why S1-right carries the
    timing instead of S2-right, but it says the MLU is
    strobing its inputs correctly.