Main INDEX
Monthly INDEX
PREV
NEXT
User name R.Michaels
Log entry time 02:07:26 on February 17,1999
Entry number 14678
Followups:
keyword=scaler cable replacement
Replaced the 34-pin flat cables that fanout the signals
for the normalization scalers (charge, triggers, etc) on
the H-spectrometer. There was fairly good evidence that
the d10 anomaly was traceable to one or more bad cables
at this fanout. The connectors can be intermittent.
Wang and I did some test runs. The sum rule between
+ and - helicity seems violated now by u1 and d1 charge !!
We see "+ " is about same as sum, and "-" is ~0.5*sum.
On E-arm, the sum rule is satisfied for these. Will investigate...