User name Jack
Log entry time 17:26:32 on February 21,1999
Entry number 15113
keyword=Gas system exhaust pressure switches.
I adjusted the pressure switches on the detector stacks. The ones for the VDCs
were set just above the value they are sitting at. Apparently the jitter in the pressure
of the exhaust line was sufficient to cross the threshold. The switches are now set
to .30 or .35 inches of water. If the alarms from the gas system occur again page
me.