Main INDEX
Monthly INDEX
PREV
NEXT
User name R. Michaels
Log entry time 22:03:02 on April 3,1999
Entry number 16999
This entry is a followup to: 16997
keyword=remarks about CODA problems
When I came in, the 3 fastbus crates were in a strange state,
and pressing green reset buttons for 10 seconds fixed it.
But the main problem this weekend has been with the hadron scaler
crate TS1/halladaq4. The scaler server answers slowly and sometimes
crashes (type "i" in the login session, see "scalServ" is "suspend").
I saw two of these crashes; they happened at a machine address
which did not correspond to the scalers but was probably in the cpu.
Anyway, when this server crashes, datamon and xscaler don't work, and
you can't run because you get no scaler data. This problem seems to
come and go (mostly it is gone). If it persists, we can try some things but
they could take 1-2 hours of beam time. Your choice.