MODEL 1880B DUAL READOUT SCALER

The Model 1880 is a dual NIM width module containing two identical channels of eight (8) digit BCD 100MHz scalers. Readout is provided on eight (8) in line LED readouts for each channel.

        SPECIFICATIONS





 CAPACITY         Eight (8) digits BCD plus overflow.



 SIGNAL INPUT(each channel)



  Pulse Repetition Rate:    0 to 100mpps min(125mpps typical)

  Pulse Pair Resolution:    10 ns min.

  Minimum Pulse Width:     3nsec FWHM

  Discriminator Level:     -500mv (internally adjustable to a

           min. -300mv)

  Input Impedance:      50Ohm 10% D.C. coupled

  Input Reflections:     Less than 10% for lns rise time of

           NIM level signals.



 GATE (each channel)



  Manual:        Front panel three position switch

           (OPEN,REMOTE,CLOSED) for controlling

           the counting gate. In the remote

           position electrical gate signals

           control the counting gate.



  Electrical:       Two bridged LEMO connectors on front

           panel. Bridged inputs allow daisy

           chaining of gate signals from scaler

           to scaler.



  Input Impedance:      5K min.

  Sensitivity:       -500mv

  Input Protection:     t 50 volts transients

  Response Time:      Less than 10nsec.



 RESET (each channel)



  Manual:        Front panel push button for reset of

           counter overflow and display.



  Electrical:       Two bridged LEMO connectors on front

           panel. Bridged inputs allow daisy

           chaining of gate signals from scaler

           to scaler. The input is biased to a

           CAMAC logic zero (approx. 3.5v) and

           a ground or a current sink of 10ma

           will reset scaler channels to zero.





 DISPLAYS (each channel)



   Counter:        Eight (8) inline 3/8" high character

            LED seven(7) segment digits.



   Gate:         LED light on when counter gate is

            open.



   Overflow:        LED light on when counter overflows.

            Turns off when counter is reset.



 OUTPUTS (each channel)



   Carry Output:       LEMO front panel connector provides

            a fast NIM pulse -800mv into 50 ohms

            each time counter overflows. A

            carry pulse will not be generated

            at reset.



 POWER REQUIREMENTS:       Less than 12 watts

            + and - 6 volts



 TEMPERATURE RANGE       0 to 60'C



 SIZE          Double width 8 3/4 high NIM module.