Click model number for photo
1877
MULTIHIT TIME-TO-DIGITAL CONVERTER
- High Density, 96 Channels Per FASTBUS Slot
- 500 psec LSB
- Programmable Full Scale up to 32 µsec
- Up to 16 Hits/Channel (Programmable)
- Rising and/or Falling Edge Timing Information
- < 20 nsec Double Pulse Resolution
- Fast Conversion
- Built-in Data Zero Suppression and Data Compaction
- Multiple Event Buffer, 7 Events
- High-Speed FASTBUS Block Transfers
- Supports the Multi-Module Data Transfer Specification (Multi-Block)
- Built-in Self Test
HIGH RATE TDCs FOR WIRE CHAMBER SPECTROMETERS
The FASTBUS Model 1877 is a 96-channel, high resolution, multihit TDC designed
for high rate experiments. This TDC provides precise time measurements for
spectrometers and hodoscopes consisting of multiwire propor tional, drift
or time projection chambers. With a least count of 500 psec and a 16-bit
range, the Model 1877 can measure time intervals up to 32 µsec duration.
The instrument operates in Common Start or Common Stop modes. Common Stop
Mode eliminates the need for expensive delay cables normally required for
each input channel of a fixed target experiment.
The 1877 provides the high performance, high density, versatility, and automated
test features required for large scale experiments. By using the FASTBUS
standard, the LeCroy 1800 family of data acquisition modules is one of the
most sophisticated data acquisition systems today.
FUNCTIONAL DESCRIPTION
Input Circuits
The individual channel inputs are located on the front panel. They accept
differential ECL signals and are termi nated in 110 ohm. The COMMON input
(Common Start or Stop) also accepts ECL signals and has a removable 110
ohm termination. By removing this termination, the COMMON signal(s) can
be cascaded, as long as the termination in the last module of a block of
modules is left in place. This scheme eliminates the extra propagation delays
caused by embedded repeaters and leaves only the very slight delays due
to the cables cascading the individual modules (approximately 50 psec/cm).
Conversion Technique
At the heart of the 1877 lies the MTD133B Monolithic Time Digitizer, an
8-channel, 16-bit dynamic range TDC circuit with 500 psec least count. Arrival
times can be recorded for either the rising, falling or both edges. The
double pulse resolution is 20 nsec.
The MTD133B uses a high-speed clock, a continuously counting scaler, and
four state interpolators. When a pulse arrives, the contents of the scaler
and interpolator are stored in memory. Up to 16 words of data can be stored
in the LIFO (Last In First Out memory) for each channel. During readout
each of the stored signal times is automatically subtracted from the COMMON
hit time. Therefore, the output is the actual time difference with virtually
no pedestal. The difference value is further compared to a programmable
upper threshold and discarded if it is greater than this threshold. This
threshold is programmable from 8 nsec to 32 µsec with a resolution
of 8 nsec.
In this way the MTD133B accommodates many different, difficult measurement
scenarios. For users who wish to detect only pulses coming within the drift
time of the chamber, the programmable threshold produces a program mable
full scale. In addition, the user can program the depth of the LIFO from
1 to 16 hits, thus creating a guaran teed maximum conversion time.
Control
Control registers configure the TDC to report either the time of the rising
edges, falling edges or both, facilitating pulse width measurements. Test
functions allow verification of each channel's behavior. Full scale time
out in the COMMON START mode can either be programmed or supplied externally.
Compatibility with the Model 1810 Calibration and Timing Module is maintained
for COMMON START/STOP, fast clear, fast clear window, and test pulse distribution.
Multiple Event Buffering
The module contains a seven event buffer. This digital memory buffer provides
two primary advantages. First, dead time in the experiment is reduced because
data readout can be done during the acquisition of subsequent events. Second,
the event data can be stored temporarily while the trigger decision to read
or discard the event is made. Events in the buffer are discarded with a
FASTBUS command to skip the event. This skip command causes an internal
pointer to increment, positioning the next event at the top of the readout
queue. As each event is recorded, a modulo eight event tag number is appended
to it in order to allow the coherence across multiple modules to be verified.
FASTBUS Readout
The 1877 complies with the FASTBUS Standard (ANSI/IEEE-960). FASTBUS functions
allow remote control and operation of the 1877. Data from the 1877 is automatically
zero suppressed. Readout is in a FIFO-like manner, consisting of a header
word followed by a variable number of data words. Channel identification
information is appended to each data word.
The modules may be read out via a LeCroy Model 1821 FASTBUS Segment Manager/Interface
(SM/I) at data transfer rates up to 10 Mwords/sec. In addition, the Model
1877 is compatible with the LIFT (LeCroy Interactive FASTBUS Toolkit) software
package.
Applications
The Model 1877, 96-input TDC, has been designed to be as compatible as possible
with the popular Model 1879 Pipelined TDC. With its shorter conversion time,
higher precision and larger dynamic range, the 1877 is the perfect replacement
for the 1879 in high rate, high accuracy applications.
By capturing both the rising and falling edge time information, the Model
1877 can be used in a "time-over -threshold" technique to simultaneously
determine the time and total charge collected or, in some cases, the time
and peak value for a given detector element using only one channel of TDC
electronics. This technique avoids, at a greatly reduced cost, many of the
traditional problems encountered with other methods, such as amplitude saturation.
Model 1877 Block Diagram
SPECIFICATIONS
Inputs: 96 ECL differential line receivers. Input impedance 110 ohm
±10%. Minimum pulse width 10 nsec FWHM (must be > 1 time bin width).
Input swing 400 mV, differential.
Least Significant Bit: 500 psec.
Total R.M.S. Error: 400 psec (Note: The R.M.S. of a gaussian distribution
is equal to sigma).
Time Out: Differential ECL input to mark the end of acquisition in
COMMON START mode.
Full Scale: 0 to 32.768 µsec, ±0.0025%; programmable via
CSR18 in steps of 8 nsec.
Pedestal: 14 to 20 counts.
Double Edge Resolution: The 1877 can measure two edges separated
by as little as 10 nsec. No two pulse edges should be closer than 10 nsec.
This implies a lower limit on the double pulse resolution of 20 nsec.
Common Start/Stop: From the Model 1810 CAT via TR6 line or from front-panel
differential ECL input. CSR selected.
Fast Clear Window Input: From the Model 1810 CAT via TR5 line.
Fast Clear Window (FCW): Starts at end of Time Range (Common Start)
or at Common Stop. Can be pro grammed 1024 nsec to 512 µsec. During
this period, the user can apply a FAST CLEAR to discard the event just captured.
Zero Suppression: Automatic for channels that have no hit.
Long Term Stability: < 100 ppm/year.
Temperature Coefficient: < 10 ppm/°C.
Differential Non-Linearity: Maximum ±0.2 LSB (typical).
Integral Non-Linearity: < 25 ppm full scale.
Fast Clear: Differential ECL input via a 2-pin front-panel connector
(removable termination resistors) or via backplane TR0 line. Minimum pulse
width 40 nsec.) When applied during the FCW, clears data in the current
event and readies module for acceptance of a new event. Fast clear settling
time is < 250 nsec. Fast Clear is leading edge sensitive and must be
performed during FCW.
Time Out: Differential ECL input via a 2-pin front-panel connector.
Minimum width 50 nsec. In Common START mode, terminates measurement in progress
and starts conversion.
Busy Output: Differential ECL output via a 2-pin front-panel connector.
Indicates the module is converting hit information. The unit is unavailable
for data capture.
Conversion Time: 750 nsec + 50 nsec per hit within the programmed
full scale; 1.6 µsec minimum.
On-Board Tester: The tester generates square wave pulses (50% duty
cycle). The pulse trains can have 1, 2, 4 or 8 cycles with half periods
of 125, 250, 1000 nsec or 2000 nsec.
Multiple Event Buffer: The digital data memory is logically organized
as a circular buffer, large enough to store the results of up to 7 events
when used appropriately. One buffer is always dedicated for FASTBUS readout.
GENERAL
Front-Panel Indicators: Slave: Indicates module is being addressed. COMMON:
Indicates whether Common Start/Stop was hit.
Power Requirements: 5 V at 5.0 A, -5.2 V at 4.0 A, -2 V at 3.0 A,
15 V at 100 mA, -15 V at 100 mA.
Packaging: Single-width FASTBUS module (ANSI/IEEE-960-1989).
FASTBUS CONTROL
Module Identification Code: Read Only, 103Ch.
Implemented Addressing Modes: Logical (16 bits), Geographical, Broadcast.
AS-AK Handshake Time: 125 nsec typical, 150 nsec maximum.
DS-DK Handshake Time: 65 nsec typical, 75 nsec maximum (Block Transfer).
Implemented Broadcast Functions:
(01)h - General Broadcast Select: The TDC modules are selected and respond
to subsequent data cycles.
(05)h - Class N Broadcast: The TDC modules of Class N (programmed via CSR7)
respond to subsequent data cycles.
(09)h - Sparse Data Scan (SDS): TDC modules containing one or more buffered
events assert their "T pin" on the following read data cycle.
(19)h - Device Available Scan (DAS): TDC modules respond by asserting "T
pin" if no events are buffered.
(0D)h - All Device Scan: All TDC modules assert their T pin on the following
read data cycle.
(BD)h - AFC SDS: All TDC modules with AFCs requiring service assert their
T pins.
(CD)h - TDC DS: TDC modules assert T pin if current event contains a non-zero
number of data words.
* An h subscript denotes a hexadecimal number, i.e., base 16.
Slave Status Responses to Data Cycles:
SS Significance
0 = Valid action.
2 = End of data.
3 = Error: Error in token pass during multi-module data scan.
7 = Error. Invalid secondary address loaded into internal address register.
DSR0 Output Word Bit Definitions
When logical addressing is enabled use top configuration.
When logical addressing is disabled use bottom configuration.
D = Data
S = Edge: Rising Edge = 0, Falling Edge = 1 (on "+" inputs)
A = Channel number 0 - 95
G = Geographic address
L = Logic address
P = Parity
W = Word Count
B = Buffer
Auxiliary Functions Card Socket - Viewed From Front of Crate (Reverse
for Rear View)
Auxiliary Connector Pin Out Description
Trigger Bits: TTL, active low signals. Front-panel signals are converted
to TTL and routed to Auxiliary Connec tor.
Trigger Strobe: A signal received by the TDC via the FASTBUS segment
from the 1810 CAT module. Normally used by the AFC to define the fiducial
time interval.
Aux_DB0* - Aux_DB11*: A 12-bit bidirectional bus. TTL, active low.
Aux RDB: Defines direction of data bus Aux_DB0-11. When high AFC
is in read mode (i.e., being read from the Segment).
Aux WCLK*: All read/write clock applied to auxiliary card whenever
the user accesses the user CSR space C0000000-C000000Fh.
Address Lines: CSR_ADR 0-3; Address lines CSR_ADR 0-3 in conjunction
with the decode AFC address strobe (Aux WCIK*) allows user implementation
of FASTBUS CSR locations C0000000 to C000000Fh. Sixteen locations are available
for use on the AFC. CSR_ADR 0-3 are latched on 1877 card.
V_Addr*: TTL active low signal to be driven by AFC circuits if an
implemented address is being accessed on the AFC. Used to generate the proper
SS = 0 response to FASTBUS; otherwise, SS = 7 is generated if Valid Address
is not driven low.
PCT*: TTL active low signal asserts module T pin in response to a
(BD) h broadcast.
FC*: TTL active low signal equal in duration to the Fast Clear input
applied via the front panel or 1810 CAT module.
Power Supply: All FASTBUS voltages.
* Indicates a low true signal.
Copyright© April 1996. LeCroy is a registered trademark of LeCroy Corporation.
All rights reserved. Information in this publicaction supersedes all earlier
versions.