INPUT CHARACTERISTICS A, B, C, D, Four inputs per section, LEMO connectors; accepts NIM level logic signals (-500 mv); 50 ohm input impedance direct coupled; input reflections are less than +/-5% for a 1 nSEC rise- time. Inputs are protected against damage from +/-50 volt input transients. Inputs respond to a 1 nSEC or greater input width. Fast Veto, One input per section, LEMO connector; accepts NIM level logic signal (-500 mV); 50 ohm input impedance, direct coupled; less than +/-5% input reflection for a 1 nSEC risetime, protected against damage +/-50 volt input transients. Requires a 3.5 nSEC minimum input width in time with the input pulse leading edge to inhibit. Bin Gate, Rear-panel slide switch enables or disables the slow bin gate via the rear connector. Signal levels are in accordance with the TID-20893 standard. INPUT CHARACTERISTICS General Five outputs per section, two pairs of negative bridged and one complemented NIM. The two pairs of bridged outputs are quiescently 0 mA and -32 mA during output. (-1.6 V into 50 ohms or -.8 V into 25 ohms). The complemented output is quiescently -16 mA and 0 mA during output. Risetimes and falltimes are less than 1.5 nSEC, and the output pulse shapes are optimized when the bridged outputs are 50 ohm terminated. Width Control One control per section; 15-tum screwdriver adjustment. Outputs are continuously variable from 4 NSEC to 1 uSEC; better than 0.15%/OC. Updating Operation: The output pulse will be extended if a new input pulse occurs while the output is active. This provides deadtimeless operation and 100% duty cycle can be achieved. GENERAL PERFORMANCE Functions: Logic AND, OR, majority logic, and logic fan-in/fan-out. All functions have leading edge inhibit with standard- ized outputs. Rate: 150 MHz minimum, input to output. Typically 160 MHz. Double-Pulse Resolution: Less than 6.5 NSEC; Typically 6 NSEC with output width set at minimum. Input to Output Delays Less than 8 NSEC. Multiple Pulsing: One and only one output pulse regardless of input pulse amplitude or duration. Power Supply Requlrements: -6 V @ 400 mA +6 V @ 250 mA -12 V @ 165 mA +12 V @ 0 mA -24 V @ 60 mA +24 V @ 35 mA 115 VAC @ 60 mA Note All currents within NIM specifications limits allowing a full-powered bin to be operated without overloading. Operating Temperature# O'C to 70'C ambient. Packaging: Standard single width NIM module in accordance with TID-20893 and Section 524. Options# Call Phillips Scientific to find out about available options.