The P6137 is a compact, 1.5 meter, 10X, passive voltage probe designed specifically for use with the TEKTRONIX 2400 Series oscilloscopes. It is fully compatible with the Tektronix family of compact pr-)be accessories.


 Electrical Characteristics (Probe installed on 2400 Series Oscilloscopes)

 Attenuation: 1 OX +/- 1 % at dc.

 Input Impedance (System): 1 0 M ohm shunted by 10.8 pF at DC (See Figure 1.)

 Compensation Range: 12 pF to 18 pF.

 Signal Delay: Delay difference between any two probes of equal length is <200 ps.

 System Bandwidth (-3dB): Use of the P6137 on a 2400 Series oscilloscope with a bandwidth measured

 to be greater than 350 MHz from a properly terminated 50 ohm source, and having an input capacitance

 of 15 pF +/- 2 pF, will result in a system bandwidth of at least 350 MHz. A similar correspondence is

 achieved for all 2400 Series oscilloscopes of identical input capacitance and different bandwidths up to

 a system maximum bandwidth of 400 MHz.

 Maximum Nondestructive Input Voltage: 500 V (dc + peak ac) to 1.3 MHz. (See Figure 2 for voltage

 derating curve.)

 System Aberrations less oscilloscope's 50-ohm-DC coupled aberrations: 4% overshoot, during first 2 ns,

 <= +/-3%, 5% p-p in the following 30 ns, +/-l %,l % p-p thereafter.







 Environmental Characteristics

 Temperature Range (Operating): - 15'C to +75'C (+5'F to +167'F).

 Temperature Range (Nonoperating): - 62'C to + 85'C (- 80'F to +185 'F).

 Humidity:Five cycles (120 hr) at 95% to 97% relative humidity. Per Tek Standard O62-2847-00,Class

 3. ReftoMIL-E-16400F,paragraph4.5.9through4.5.9.5.1,class4.

 Physical Characteristics

 Net Weight (includes Accessories): 168 g (6 oz).

 Probe Cable Length: (1.5 m ~= 5 ft).