It was noticed that the efficiency of the FPP chambers had drop to a low
level. This was indicated by a very low number of hits in the "fpp_nhit"
spectra. Normally this spectra peaks at about 23 hits but was peaking at
4 hits. Other spectra ( TDC, Demux, wire maps) looked ok in shape with
just the number of hits per 10000 triggers lower than usual when compared
to the "golden" run.
It seems that the gas mixture for the FPP is the culprit. They
found ( Rob, Alexandre, Mark Stevens) that the argon gas
had not switch to a new bottle. Why this effected the FPP and not the VDC
is a question. But eveything else with the FPP seems fine. The gas flow,
HV , LV and theshold are ok. I went into the hall and verified the
gas flow and LV setting by eye.
I would guess that it would take at least 6 hours for the FPP
to recover. I have set up extra histograms in the online replay
to monitor the number of hits per chamber. I suggested to the
shift crew that they monitor it.