Difference between revisions of "Cryo-target wall contamination"

From Hall A Wiki
Jump to: navigation, search
(Kinematic 6.5)
Line 72: Line 72:
  
 
Result: An estimate of 4439 events wall events remains within the cut ==> Contamination is 4%.
 
Result: An estimate of 4439 events wall events remains within the cut ==> Contamination is 4%.
 
  
  
Line 78: Line 77:
  
 
[[Image:dummy_6.5_all.gif]]
 
[[Image:dummy_6.5_all.gif]]
 +
 +
 +
* With the full statistic collected till run 4039:
 +
 +
[[Image:dummycont_6.5.gif]]

Revision as of 14:48, 1 May 2011

The normalized yield is equal to Ngood/Q/density. Here I neglected the density of the 3He gas.


Kinematic 5.2

Target window thickness Total charge # of good evts
upstream downstream
He3 0.0274 cm 0.0328 cm 6.92E+8 253254
Dummy 0.1581 cm 0.1585 cm 2.16E+8 3588

Result: An estimate of 2186 events wall events remains within the cut ==> Contamination is less than 1%.


He3 5.2 all.gif

Dummy 5.2 all.gif


Kinematic 6.5

Target window thickness Total charge # of good evts
upstream downstream
He3 0.0274 cm 0.0328 cm 1.045E+9 110050
Dummy 0.1581 cm 0.1585 cm 1.316E+7 294

Result: An estimate of 4439 events wall events remains within the cut ==> Contamination is 4%.


He3 6.5 all.gif

Dummy 6.5 all.gif


  • With the full statistic collected till run 4039:

Dummycont 6.5.gif