Don't forget to make new eventfile
 +
 +
===Assumptions:===
 +
*Fridge running − all pumps on
 +
Prepare NMR:
 +
*Stop beam, if necessary
 +
*Turn off Microwaves, if necessary
 +
*Put NMR into Monitor Mode
 +
===Prepare Fridge:===
 +
*Stop Roots Blower 3 by pressing the RB3 Stop Button (in electronics room)
 +
*  Wait 2 minutes for pump to spin down
 +
*  Stop Roots Blower 2 by pressing the RB2 Stop Button (in electronics room)
 +
*  Wait 2 minutes for pump to spin down
 +
*  Stop Roots Blower 1 by pressing the RB1 Stop Button (in electronics room)
 +
*  Open Main Gate Valve, PV91141, if necessary (in electronics room)
 +
*  Close Bypass RB3 Valve, PV91142, if necessary (in electronics room)
 +
*  Close Roughing Valve, PV91143, if necessary (in electronics room)
 +
*  Place Run Valve, EV91120, into Manual Mode (cryo computer)
 +
*  Close Run Valve by entering a manual setpoint of zero
 +
*  Close Bypass Valve, EV91121, if necessary, by entering a position of zero
 +
*  Put the Separator Valve, EV91127, into Computer Control (not Manual Mode)
 +
*  Enter a value of 60 into the Set Val box of the EV91127 control
 +
===Empty the Tail of Helium:===
 +
*  DO NOT move the target without first informing MCC − youʼ’ll trip all Halls
 +
*  Move the target to the Top position, write in logbook
 +
*  Load the Anneal program (icon on desktop)
 +
*  Run the Anneal program (click white arrow on left of toolbar)
 +
*  Type in a setpoint of 60 (K) and hit "Send to ITC", write in logbook
 +
*  Hit the "Goto Setpoint" button to turn on the heater
 +
*  Observe the liquid level in the tail drop (7% is about the minumum reading)
 +
*  Wait 5 minutes after the liquid is gone
 +
*  Open the Run Valve to 0.3, write in logbook
 +
*  Move the target to Empty position, write in logbook
 +
*  (If Run Plan needs to do Carbon runs, this position is also OK)
 +
*  Use Lower camera to see the He4 pressure (Rack B, Device 5), write in logbook
 +
===Begin the Anneal:===
 +
*  Wait until all three sensors stabilize at 60K, write in logbook
 +
*  Type the desired Anneal temperature into the setpoint, Hit "Send to ITC"
 +
*  Note in the logbook the time when the anneal temperature is reached
 +
*  Log Top Platinum, Top T/C, Bottom T/C, and He4 Pressure every 5−10 minutes
 +
*  Leave the target at the Anneal temperature for the desired number of minutes
 +
*  To stop the anneal, hit the "Stop Anneal" button, write in logbook
 +
*  Let the anneal program continue to run, to document the cooldown process
 +
*Cool Down the Refrigerator:
 +
*  Change the setpoint of the Bypass Valve to 1.0
 +
*  Change the Manual setpoint of the Run Valve to 1.0
 +
*  Wait until the Nose Level, LL91112, reaches about 80%
 +
*  Change the setpoint of the Bypass Valve to 0.0
 +
*  Change the Run Valve back to computer control (not Manual Mode)
 +
*  Enter a value of 32 into the Set Val box of the EV91127 (Separator) control
 +
*  The the Stop button on the toolbar of the Anneal Program, and then close it
 +
*  Wait for the Nose Level to (mostly) stabilize
 +
*  Observe the He4 pressure
 +
*  If the pressure is not below 12 torr, temporarily close the Run Valve
 +
*  Once the pressure is below 12 torr, start RB1 (electronics room)
 +
*  Wait for the pressure to drop below 2.2 torr
 +
*  Start RB2
 +
*  Wait for the pressure to drop below 1.0 torr
 +
* Start RB3
 +
*  If necessary, re−open the Run Valve or put it under computer control

Latest revision as of 11:01, 12 September 2011

g2p Anneal Procedure will go here Don't forget to make new eventfile

Assumptions:

  • Fridge running − all pumps on

Prepare NMR:

  • Stop beam, if necessary
  • Turn off Microwaves, if necessary
  • Put NMR into Monitor Mode

Prepare Fridge:

  • Stop Roots Blower 3 by pressing the RB3 Stop Button (in electronics room)
  • Wait 2 minutes for pump to spin down
  • Stop Roots Blower 2 by pressing the RB2 Stop Button (in electronics room)
  • Wait 2 minutes for pump to spin down
  • Stop Roots Blower 1 by pressing the RB1 Stop Button (in electronics room)
  • Open Main Gate Valve, PV91141, if necessary (in electronics room)
  • Close Bypass RB3 Valve, PV91142, if necessary (in electronics room)
  • Close Roughing Valve, PV91143, if necessary (in electronics room)
  • Place Run Valve, EV91120, into Manual Mode (cryo computer)
  • Close Run Valve by entering a manual setpoint of zero
  • Close Bypass Valve, EV91121, if necessary, by entering a position of zero
  • Put the Separator Valve, EV91127, into Computer Control (not Manual Mode)
  • Enter a value of 60 into the Set Val box of the EV91127 control

Empty the Tail of Helium:

  • DO NOT move the target without first informing MCC − youʼ’ll trip all Halls
  • Move the target to the Top position, write in logbook
  • Load the Anneal program (icon on desktop)
  • Run the Anneal program (click white arrow on left of toolbar)
  • Type in a setpoint of 60 (K) and hit "Send to ITC", write in logbook
  • Hit the "Goto Setpoint" button to turn on the heater
  • Observe the liquid level in the tail drop (7% is about the minumum reading)
  • Wait 5 minutes after the liquid is gone
  • Open the Run Valve to 0.3, write in logbook
  • Move the target to Empty position, write in logbook
  • (If Run Plan needs to do Carbon runs, this position is also OK)
  • Use Lower camera to see the He4 pressure (Rack B, Device 5), write in logbook

Begin the Anneal:

  • Wait until all three sensors stabilize at 60K, write in logbook
  • Type the desired Anneal temperature into the setpoint, Hit "Send to ITC"
  • Note in the logbook the time when the anneal temperature is reached
  • Log Top Platinum, Top T/C, Bottom T/C, and He4 Pressure every 5−10 minutes
  • Leave the target at the Anneal temperature for the desired number of minutes
  • To stop the anneal, hit the "Stop Anneal" button, write in logbook
  • Let the anneal program continue to run, to document the cooldown process
  • Cool Down the Refrigerator:
  • Change the setpoint of the Bypass Valve to 1.0
  • Change the Manual setpoint of the Run Valve to 1.0
  • Wait until the Nose Level, LL91112, reaches about 80%
  • Change the setpoint of the Bypass Valve to 0.0
  • Change the Run Valve back to computer control (not Manual Mode)
  • Enter a value of 32 into the Set Val box of the EV91127 (Separator) control
  • The the Stop button on the toolbar of the Anneal Program, and then close it
  • Wait for the Nose Level to (mostly) stabilize
  • Observe the He4 pressure
  • If the pressure is not below 12 torr, temporarily close the Run Valve
  • Once the pressure is below 12 torr, start RB1 (electronics room)
  • Wait for the pressure to drop below 2.2 torr
  • Start RB2
  • Wait for the pressure to drop below 1.0 torr
  • Start RB3
  • If necessary, re−open the Run Valve or put it under computer control