* Run list
+
* Initial studies (Oct 28, 2011)
** General Setup: 90Sr source with scintillator below the GEM as the trigger. Threshold on the scintillator is 40mV. The GEM is a "pad" readout. The gas for GEM is Ar/CO2 at 75/25 mixture.  
+
** General Setup: 90Sr source with scintillator below the GEM as the trigger. Threshold on the scintillator is 40mV. The GEM is a "pad" readout with one preamp. The gas for GEM is Ar/CO2 at 75/25 mixture.  
 
** DAQ is using the flash ADC with GEM in Chan 1 and Scintillator in Chan 2.
 
** DAQ is using the flash ADC with GEM in Chan 1 and Scintillator in Chan 2.
 
{| class="wikitable" style=text-align: center" cellspacing = "0" border = "1"
 
{| class="wikitable" style=text-align: center" cellspacing = "0" border = "1"

Revision as of 10:52, 28 October 2011

Studies for GEM "pad" readout

  • Initial studies (Oct 28, 2011)
    • General Setup: 90Sr source with scintillator below the GEM as the trigger. Threshold on the scintillator is 40mV. The GEM is a "pad" readout with one preamp. The gas for GEM is Ar/CO2 at 75/25 mixture.
    • DAQ is using the flash ADC with GEM in Chan 1 and Scintillator in Chan 2.
Run Number GEM HV Comment
1025 3.4kV Low voltage on preamp = 3.5V
1026 3.52 Scint signal was missing. Forgot to plug it back in after looking at scope.
1027 Event rate = 56/sec
1028 No source. Event rate = 10.4 /sec. So maximum GEM efficiency is 81.4%.
1029
1030
1031
1032 Low voltage on preamp = 5.0V
1033