Difference between revisions of "How to take HARP scan"

From Hall A Wiki
Jump to: navigation, search
Line 9: Line 9:
 
[[Image:Harp1H05.png]]
 
[[Image:Harp1H05.png]]
 
*In Monticello, go in second HARP menu ( ! ) choose HARP analyzer  
 
*In Monticello, go in second HARP menu ( ! ) choose HARP analyzer  
 +
[[Image:Harp-analyzer.png]]
 
*Select the Hall A harp to use
 
*Select the Hall A harp to use
 
*Click on Initiate scan
 
*Click on Initiate scan
 
*If signals saturates, reduce gain on the corresponding HARP and take another scan
 
*If signals saturates, reduce gain on the corresponding HARP and take another scan
* Once done, go to 50 nA, take a run
+
* Once done, go to 50 nA, take a run, with ps8=1

Revision as of 18:23, 9 April 2012

After first initial tune, we want at least one HARP scan on 1HA05A to determine absolute beam position after chicane change.

  • Request 8 uA tune beam
  • In Monticello, go in first HARP menu, select HARP expert controls and open panels for 1H05A and 1H04 ( bottom right of expert screen )

Monticello.png

  • Make sure speed is set to 1

1HA1H04.png

Harp1H05.png

  • In Monticello, go in second HARP menu ( ! ) choose HARP analyzer

Harp-analyzer.png

  • Select the Hall A harp to use
  • Click on Initiate scan
  • If signals saturates, reduce gain on the corresponding HARP and take another scan
  • Once done, go to 50 nA, take a run, with ps8=1