Difference between revisions of "Yerevan bars - Test"

From Hall A Wiki
Jump to: navigation, search
Line 5: Line 5:
 
Y04 was bleached using the new lamp and Y05 with the old one, in different sessions for an accumulated time of 10h.
 
Y04 was bleached using the new lamp and Y05 with the old one, in different sessions for an accumulated time of 10h.
  
 
+
The bleaching process shows a particular behavior in the first centimeters. It shows that there is a quite different bleaching time evolution within the bar, caused, perhaps, by the different damage level and high absorption in the first section of the bar. Anyhow, this is not clearly understand.
 
+
 
+
 
+
  
  

Revision as of 22:37, 20 July 2014

Two bars (Y04 and Y05) from Yerevan Institute in Armenia, previously used in BigCal of GEp(3) were tested making use of the new lamps and the lamps used in GEp(3). This was a first approach to compare how good is the improvement of the new lamps.

The bars have a dimensions of 4.0 x 4.0 x 40 <math>cm^3</math>. Y04 was irradiated at the RadCon facilities in JLab under <math>^{137}Cs</math> for ~70h. Y05 was irradiated for 68h. The dose rate is 280Rad/h.

Y04 was bleached using the new lamp and Y05 with the old one, in different sessions for an accumulated time of 10h.

The bleaching process shows a particular behavior in the first centimeters. It shows that there is a quite different bleaching time evolution within the bar, caused, perhaps, by the different damage level and high absorption in the first section of the bar. Anyhow, this is not clearly understand.


Measurement Y04 (bleached with new lamp) Y05 (bleached with old lamp)
Transversal Transmission vs wavelength
Fig.5 ID06 bar, transmission vs depth. 8h accumulated UV sessions
Fig.5 ID06 bar, transmission vs depth. 8h accumulated UV sessions
Transversal Transmission vs depth for six different wavelengths
Fig.5 ID06 bar, transmission vs depth. 8h accumulated UV sessions
Fig.5 ID06 bar, transmission vs depth. 8h accumulated UV sessions
Transversal Transmission (at 0.5cm) vs time for 400nm
Fig.5 ID06 bar, transmission vs depth. 8h accumulated UV sessions
Fig.5 ID06 bar, transmission vs depth. 8h accumulated UV sessions
Transversal Transmission (at 2.0cm) vs time for 400nm
Fig.5 ID06 bar, transmission vs depth. 8h accumulated UV sessions
Fig.5 ID06 bar, transmission vs depth. 8h accumulated UV sessions
bottom-left-cell bottom-center-cell bottom-right-cell