How to perform a BPM pedestal run:
 
How to perform a BPM pedestal run:
  
1.   MCC hast to follow the following procedure
+
1. MCC hast to follow the following procedure:
      - open BPM window "BPM Diagnostics - SEE"
+
    - open BPM window "BPM Diagnostics - SEE"
      - from there pull down the "Expert Screens" and open "Gain Control"
+
    - from there pull down the "Expert Screens" and open "Gain Control"
      - for IOCSE10, use the pull down window and change from "auto gain" to "forced gain". Then change the Forced Gain values x and y to zero.
+
    - for IOCSE10, use the pull down window and change from "auto gain" to "forced gain". Then change the Forced Gain values x and y to zero.
  
2.   CODA has to run in pedestal mode
+
2. CODA has to run in pedestal mode:
      - if you were running before you first have to press Abort
+
    - if you were running before you first have to press Abort
      - choose the appropriate configuration in runcontrol
+
    - choose the appropriate configuration in runcontrol
      - select "PEDRUN"  - Press RunTypes button in RunControl
+
    - select "PEDRUN"  - Press RunTypes button in RunControl
 +
 
 +
3.  Run get_bpm_pedestals.C in the root analyzer:
 +
    - you might have to modify some details
 +
    - the code should print out the new pedestals at the end
 +
    - load the database into an editor and change them (2nd line of the [BPMx] blocks)
  
 
== Moller polarimeter and results of measurements ==
 
== Moller polarimeter and results of measurements ==

Revision as of 10:45, 30 September 2008

E06-010 Instrumentation Requirement List: Jian-Ping Chen's memo to Kees De Jager, 08/07/08

.doc file and htm file

ARC energy measurement

Compton Polarimeter

HARP and HARP scan procedures

Please follow the directions at the Harp Scan Page [Harp Scan page]

Beam raster

Beam charge monitors (BCM) and calibration procedure

Beam position monitors (BPM) and calibration procedure

How to perform a BPM pedestal run:

1. MCC hast to follow the following procedure:

    - open BPM window "BPM Diagnostics - SEE"
    - from there pull down the "Expert Screens" and open "Gain Control"
    - for IOCSE10, use the pull down window and change from "auto gain" to "forced gain". Then change the Forced Gain values x and y to zero.

2. CODA has to run in pedestal mode:

    - if you were running before you first have to press Abort
    - choose the appropriate configuration in runcontrol
    - select "PEDRUN"  - Press RunTypes button in RunControl

3. Run get_bpm_pedestals.C in the root analyzer:

    - you might have to modify some details 
    - the code should print out the new pedestals at the end 
    - load the database into an editor and change them (2nd line of the [BPMx] blocks)