== Beam position monitors (BPM) and calibration procedure == | == Beam position monitors (BPM) and calibration procedure == | ||
− | How to perform a BPM pedestal run: | + | <FONT Color="Green">How to perform a BPM pedestal run</FONT>: |
1. MCC hast to follow the following procedure: | 1. MCC hast to follow the following procedure: | ||
Line 41: | Line 41: | ||
- load the database into an editor and change them (2nd line of the [BPMx] blocks) | - load the database into an editor and change them (2nd line of the [BPMx] blocks) | ||
− | How to perform a bulls eye scan: | + | <FONT Color="Green">How to perform a bulls eye scan</FONT>: |
1. You need unrastered beam: | 1. You need unrastered beam: | ||
− | - caution you should not do this with a target requiring rastered beam | + | - <FONT Color="red">caution you should not do this with a target requiring rastered beam</FONT> |
- use carbon, BeO, optics, or in the worst case empty instead | - use carbon, BeO, optics, or in the worst case empty instead | ||
Revision as of 17:33, 2 October 2008
Contents
- 1 E06-010 Instrumentation Requirement List: Jian-Ping Chen's memo to Kees De Jager, 08/07/08
- 2 ARC energy measurement
- 3 Compton Polarimeter
- 4 HARP and HARP scan procedures
- 5 Beam raster
- 6 Beam charge monitors (BCM) and calibration procedure
- 7 Beam position monitors (BPM) and calibration procedure
- 8 Moller polarimeter and results of measurements
- 9 Scattering chamber upstream and downstream beam pipe size and window thickness
- 10 HAPPEX Luminosity (Lumi) monitors
- 11 Beam dump
E06-010 Instrumentation Requirement List: Jian-Ping Chen's memo to Kees De Jager, 08/07/08
ARC energy measurement
Compton Polarimeter
HARP and HARP scan procedures
Please follow the directions at the Harp Scan Page [Harp Scan page]
Beam raster
Beam charge monitors (BCM) and calibration procedure
Beam position monitors (BPM) and calibration procedure
How to perform a BPM pedestal run:
1. MCC hast to follow the following procedure:
- open BPM window "BPM Diagnostics - SEE" - from there pull down the "Expert Screens" and open "Gain Control" - for IOCSE10, use the pull down window and change from "auto gain" to "forced gain". - then change the forced gain values x and y to zero.
2. CODA has to run in pedestal mode:
- if you were running before you first have to press Abort - choose the appropriate configuration in runcontrol - select "PEDRUN" - Press RunTypes button in RunControl
3. Have MCC change the BPM gain back to "auto gain".
4. Run get_bpm_pedestals.C in the root analyzer:
- you might have to modify some details - the code should print out the new pedestals at the end - load the database into an editor and change them (2nd line of the [BPMx] blocks)
How to perform a bulls eye scan:
1. You need unrastered beam:
- caution you should not do this with a target requiring rastered beam - use carbon, BeO, optics, or in the worst case empty instead
2. Ask MCC to steer the beam to the nominal center of the target.
3. Wait until beam is stable and perform a harp scan and a coda run during the same time.
4. Ask MCC then to steer the beam to positions around the nominal center:
- covering at least the area the raster will cover - repeat harp and coda runs