== RICH ON-LINE MONITOR SHIFT WORKERS ==
 
== RICH ON-LINE MONITOR SHIFT WORKERS ==
  
Go on the HallA Tools, "MEDM (HLAMain)", click on "RICH Detector" then check:
+
Go on the ''HallA Tools, "MEDM (HLAMain)'', click on <u>RICH Detector</u> then check:
 
  * '''Radiator Pressure''' ~ '''10 V''',  
 
  * '''Radiator Pressure''' ~ '''10 V''',  
 
  * '''Power Supplies''' ~ '''3.8 V''' (both polarities: A nearly 3.76 V, B nearly 3.93 V),  
 
  * '''Power Supplies''' ~ '''3.8 V''' (both polarities: A nearly 3.76 V, B nearly 3.93 V),  
  
then click on "CAEN" buttom, on the "Crate 1" window popping up, click on "Modules" and then again on the next window "Crate #1" with the four modules of Crate #1, click on "Modules" of "Module 2", the only active module, then you can check and modify HV.
+
then click on the <u>CAEN</u> buttom;
 +
 
 +
the ''Crate 1'' window pop up, click on <u>Modules</u> and then again on the next window ''Crate #1'' click on <u>Modules</u> of ''Module 2'' (the only active module, then you can check and modify HV).
 
   
 
   
 
Channel 16 and Channel 17 are the only channels connected to the detector.  
 
Channel 16 and Channel 17 are the only channels connected to the detector.  
  
They have to be on Status "Normal", "HVon" have to be "ON" and the voltage "Vmon" has to correspond to the voltage "Vset", nominally between 2200 and 2300 for Channel 16 and between 200 and 400 on Channel 17.  
+
They have to be on:
 +
* Status "Normal",  
 +
* HVon have to be "ON"
 +
* the voltage "Vmon" has to correspond to the voltage "Vset" (nominally between 2200 and 2300 for Channel 16 and between 200 and 400 on Channel 17).  
  
If not, click on the "Module 2 Controls Screen" button, on the bottom of the panel, and click on "HVon" ON, then the HV should raise up very slowly, 5 V/s.
+
If not, click on the <u>Module 2 Controls Screen</u> button, on the bottom of the panel, and click on <u>HVon ON</u>, then the HV should raise up very slowly, 5 V/s.
  
If a trip occurred, the Status of the channel will be "Trip", after turning on the HV go back to the first "Crate 1" window and click on "Clear All Alarm".
+
If a trip occurred, the Status of the channel will be ''Trip'', after turning on the HV go back to the first ''Crate 1'' window and click on <u>Clear All Alarm</u>.
  
Then the status should change into "Up" and eventually into "Normal".
+
Then the status should change into ''Up'' and eventually into ''Normal''.
  
 
== Turn HV OFF DURING BEAM STUDIES OR CONFIGURATION CHANGES OF L-HRS. ==
 
== Turn HV OFF DURING BEAM STUDIES OR CONFIGURATION CHANGES OF L-HRS. ==

Revision as of 19:25, 11 November 2008

HallA RICH, old page

RICH ON-LINE MONITOR SHIFT WORKERS

Go on the HallA Tools, "MEDM (HLAMain), click on RICH Detector then check:

* Radiator Pressure ~ 10 V, 
* Power Supplies ~ 3.8 V (both polarities: A nearly 3.76 V, B nearly 3.93 V), 

then click on the CAEN buttom;

the Crate 1 window pop up, click on Modules and then again on the next window Crate #1 click on Modules of Module 2 (the only active module, then you can check and modify HV).

Channel 16 and Channel 17 are the only channels connected to the detector.

They have to be on:

* Status "Normal", 
* HVon have to be "ON"
* the voltage "Vmon" has to correspond to the voltage "Vset" (nominally between 2200 and 2300 for Channel 16 and between 200 and 400 on Channel 17). 

If not, click on the Module 2 Controls Screen button, on the bottom of the panel, and click on HVon ON, then the HV should raise up very slowly, 5 V/s.

If a trip occurred, the Status of the channel will be Trip, after turning on the HV go back to the first Crate 1 window and click on Clear All Alarm.

Then the status should change into Up and eventually into Normal.

Turn HV OFF DURING BEAM STUDIES OR CONFIGURATION CHANGES OF L-HRS.

In case of repetitive trips, keep HV OFF also during configuration changes of BigBite.

If the RICH is continuing tripping, keep HV off and call one of the RICH experts,

Do not try to keep HV on more than twice consecutively.