Difference between revisions of "IOC reboot"
From Hall A Wiki
(→Set exit line QWP) |
|||
Line 1: | Line 1: | ||
− | IOC is iochacp - might need to reboot from the hall. | + | IOC is iochacp - might need to reboot from the hall. |
− | + | No longer connected to terminal server. Can only reboot using remote power strip. | |
− | + | hareboot4. Outlet 4 | |
Before rebooting, make sure the cavity locking/unlocking script is OFF. | Before rebooting, make sure the cavity locking/unlocking script is OFF. |
Revision as of 11:46, 25 July 2019
IOC is iochacp - might need to reboot from the hall.
No longer connected to terminal server. Can only reboot using remote power strip.
hareboot4. Outlet 4
Before rebooting, make sure the cavity locking/unlocking script is OFF.
After rebooting, the quarter wave plate and half wave plate will probably be set incorrectly. Also, some power meter calibration factors need to be updated and the feedback parameters updated.
Contents
Power meter calibrations
This needs updating - commands below are absolute.
- caput COMPTON_PW3R_S2_ca.B 7.9264
- caput COMPTON_PW2R_S2_ca.B 11.833
- caput COMPTON_LASERPCAV_ca.A 13630
Feedback parameters
- caput COMPTON_SERVO_GAINao 135
- caput COMPTON_SERVO_SEUILao 1
Slow Ramp Parameters
- caput COMPTON_SLOWRAMP_AMPLITUDE 0.02
- caput COMPTON_SLOWRAMP_PERIOD 180.0
Initialize the QWP and HWP
Needs updating
- Send wave plates to 0 degrees
- caput COMPTON_ORX_bo 1
- caput COMPTON_ORZ_bo 1
- wait 2 minutes
- caput COMPTON_PVAL_X_ao 11909
- caput COMPTON_PVAL_Z_ao 5725
- caput COMPTON_LMOVX_bo 1
- caput COMPTON_LMOVZ_bo 1
This should set the QWP angle to 47.63 degrees and the HWP to 22.9 degrees
Set exit line QWP
This needs updating.
- Send QWP HOME
- caput COMPTON_ORY_bo 1
- Wait 2 minutes
- Set angle to approximately equalize power in integrating spheres
- caput COMPTON_PVAL_Y_ao 11162
- caput COMPTON_RMOVY_bo 1