Difference between revisions of "HV and Threshold calibration"
From Hall A Wiki
(New page: Procedure for the neutron array HV and threshold calibration Main considerations - * calibrate all neutron bars to equal gains. * set threshold voltage as low as posibble (50 mV), to re...) |
|||
Line 6: | Line 6: | ||
* set threshold voltage as low as posibble (50 mV), to reduce current on PMT's. | * set threshold voltage as low as posibble (50 mV), to reduce current on PMT's. | ||
* set threshold energy as low as possible to increase efficiency. | * set threshold energy as low as possible to increase efficiency. | ||
− | * set threshold energy low enough to | + | * set threshold energy low enough to remove minimum ionizing electrons. |
− | + | optimum was chosen at threshold 5 MeV - still need to verify this. | |
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
Calibrations steps - | Calibrations steps - | ||
− | * | + | * Test-area calibration using cosmic muons |
− | Use top-buttom bar trigger, to obtain | + | Use top-buttom bar trigger, to obtain vertical minimum ionizing muons with known energy deposit. |
− | + | adjust HV, so that the muon peak will be at- | |
10.0 cm bar -> E=20 MeV -> channel=1500 | 10.0 cm bar -> E=20 MeV -> channel=1500 | ||
12.5 cm bar -> E=25 MeV -> channel=1875 | 12.5 cm bar -> E=25 MeV -> channel=1875 | ||
15.0 cm bar -> E=30 MeV -> channel=2250 | 15.0 cm bar -> E=30 MeV -> channel=2250 | ||
25.0 cm bar -> E=50 MeV -> channel=3750 | 25.0 cm bar -> E=50 MeV -> channel=3750 | ||
+ | |||
+ | * Beam calibration using the H(e,e'p) reaction. | ||
+ | |||
+ | Use the H(e,e'p) reaction to create 1 GeV/c protons which will have energy loss of 32 MeV for a 10cm wide bar, calibrate the gain, so these events will be at channel=2000 | ||
+ | |||
+ | * Gain calibration cross check using the C(e,e'p) reaction, by indentifying the punch-through point of the protons. | ||
+ | |||
+ | The punch-through energy in a 10cm wide bar is 120 MeV. | ||
+ | The TOF vs. ADC of the C(e,e'p) events should have the following shape - |
Revision as of 08:03, 13 April 2010
Procedure for the neutron array HV and threshold calibration
Main considerations -
- calibrate all neutron bars to equal gains.
- set threshold voltage as low as posibble (50 mV), to reduce current on PMT's.
- set threshold energy as low as possible to increase efficiency.
- set threshold energy low enough to remove minimum ionizing electrons.
optimum was chosen at threshold 5 MeV - still need to verify this.
Calibrations steps -
- Test-area calibration using cosmic muons
Use top-buttom bar trigger, to obtain vertical minimum ionizing muons with known energy deposit.
adjust HV, so that the muon peak will be at- 10.0 cm bar -> E=20 MeV -> channel=1500 12.5 cm bar -> E=25 MeV -> channel=1875 15.0 cm bar -> E=30 MeV -> channel=2250 25.0 cm bar -> E=50 MeV -> channel=3750
- Beam calibration using the H(e,e'p) reaction.
Use the H(e,e'p) reaction to create 1 GeV/c protons which will have energy loss of 32 MeV for a 10cm wide bar, calibrate the gain, so these events will be at channel=2000
- Gain calibration cross check using the C(e,e'p) reaction, by indentifying the punch-through point of the protons.
The punch-through energy in a 10cm wide bar is 120 MeV. The TOF vs. ADC of the C(e,e'p) events should have the following shape -