Difference between revisions of "Channel Problems"
From Hall A Wiki
(4 intermediate revisions by the same user not shown) | |||
Line 7: | Line 7: | ||
[[Image:Run 29 Problem.png]] | [[Image:Run 29 Problem.png]] | ||
− | Running a source near the center of the detector reveals that the X direction is not devoid of readings | + | |
+ | |||
+ | |||
+ | Running a source near the center of the detector reveals that the X direction is not devoid of readings. | ||
+ | |||
+ | [[Image:Run_28_Source.png]] | ||
+ | |||
+ | |||
+ | Running a source near the left in the Y direction reveals it is also capable of taking readings, though very reduced compared to the X. | ||
+ | |||
+ | |||
+ | [[Image:Run_25_Source.png]] | ||
+ | |||
+ | |||
+ | |||
+ | * June 14 | ||
+ | **By trying different configurations we appear to have isolated the problem. | ||
+ | [[Image:kapton.jpg]] | ||
+ | |||
+ | ** This leads us to believe there is a problem with the soldering on the kapton. |
Latest revision as of 15:45, 14 June 2010
It appears there are problems with the middle channels on the X direction and the left channels on the Y direction. These appear to be directly related to the cards used to take their input. Switching out the X cards did not show a difference in results.
Running a source near the center of the detector reveals that the X direction is not devoid of readings.
Running a source near the left in the Y direction reveals it is also capable of taking readings, though very reduced compared to the X.
- June 14
- By trying different configurations we appear to have isolated the problem.
- This leads us to believe there is a problem with the soldering on the kapton.