== Target Holding Field == | == Target Holding Field == | ||
− | * Field survey | + | * Field survey: field strengh, gradient |
− | * Field rotation check | + | * Field rotation glitch check |
== Cell Characterization == | == Cell Characterization == | ||
Line 8: | Line 8: | ||
* Filling density | * Filling density | ||
* Broadening of Alkali absorption | * Broadening of Alkali absorption | ||
+ | * N2 density: filling density, pressure curve | ||
=== Wall Thickness === | === Wall Thickness === | ||
* Laser interference measurement: wall and window | * Laser interference measurement: wall and window | ||
+ | |||
+ | === Cell Volume/Geometry === | ||
+ | * Caliper method | ||
+ | * Archimedes' method | ||
== Optical Pumping System == | == Optical Pumping System == | ||
Line 29: | Line 34: | ||
=== NMR === | === NMR === | ||
− | * | + | * Q-curve measurement of all pickup coils to prevent resonance |
* Minimize pickup coil background with maximum RF power | * Minimize pickup coil background with maximum RF power | ||
* Linearity check of amplifiers: preamplifier, lock-in amplifier | * Linearity check of amplifiers: preamplifier, lock-in amplifier |
Latest revision as of 13:03, 9 January 2012
Contents
Target Holding Field
- Field survey: field strengh, gradient
- Field rotation glitch check
Cell Characterization
Density
- Filling density
- Broadening of Alkali absorption
- N2 density: filling density, pressure curve
Wall Thickness
- Laser interference measurement: wall and window
Cell Volume/Geometry
- Caliper method
- Archimedes' method
Optical Pumping System
Laser Optics
- Laser power measurement
- Laser spectrum measurement/Absorption adjustment
- Calibrations of optics orientations, e.g. λ/4 plates
- Laser polarization measurement
- Optics alignment [1]
- Power consumption at laser connectors (if there is any)
Oven
- RTD calibration
- Condition optimization: oven temperature as a function of gas flow and heater power
Polarimetry
NMR
- Q-curve measurement of all pickup coils to prevent resonance
- Minimize pickup coil background with maximum RF power
- Linearity check of amplifiers: preamplifier, lock-in amplifier
- Flux calibration: pickup coil survey and flux calculation
- He3 NMRs: frequency and field sweeps (including NMRs during EPR calibration)
- AFP loss measurement
- Internal temperature measurement in pumping chamber using NMR
- NMR cross calibration: frequency vs. field, different locations
- Water NMR calibration: field sweep (frequency sweep as well if possible)
EPR
- Characterization of feedback system: proportional and integration gains
- D2 light detection optimization [2]
- EPR calibration: absolute polarization measurement
- Cross check with water calibrated NMRs
- D1 light EPR measurement (if using narrow-band laser)
MISC Studies
- Spin up time
- Relaxation time: room temperature, high temperature
- Optimization of pumping condition: laser power, oven temperature
- Polarization gradient across cell