Difference between revisions of "Right Quad Tuning 2pass 48d"
From Hall A Wiki
(→Procedure without Sieve-slit) |
(→Procedure without Sieve-slit) |
||
(4 intermediate revisions by the same user not shown) | |||
Line 1: | Line 1: | ||
==== Procedure without Sieve-slit ==== | ==== Procedure without Sieve-slit ==== | ||
* <b>Kinematics need to be updated!</b> | * <b>Kinematics need to be updated!</b> | ||
− | * Time estimate: 2 hours | + | * Time estimate: 1-2 hours |
* Take a run at the following kinematics without the sieve-slit collimator. | * Take a run at the following kinematics without the sieve-slit collimator. | ||
* Start with initial right HRS spectrometer tune with the new quad in place of right Q1. | * Start with initial right HRS spectrometer tune with the new quad in place of right Q1. | ||
* If you need to increase the spectrometer momentum setting, make sure you cycle Q2 and Q3 as per the cycling procedure. | * If you need to increase the spectrometer momentum setting, make sure you cycle Q2 and Q3 as per the cycling procedure. | ||
− | * Electron rates assume 11 carbon foils, 5 mSr acceptance, and 8% delta acceptance. | + | * Electron rates assume 9 out of 11 carbon foils, 5 mSr acceptance, and 8% delta acceptance. |
* <b>A GMp expert will check the data quality and decide, if the right HRS tune needs to be adjusted to change the size of the acceptance.</b> | * <b>A GMp expert will check the data quality and decide, if the right HRS tune needs to be adjusted to change the size of the acceptance.</b> | ||
* This process will be repeated until the spectrometer tune is satisfactory.<br> | * This process will be repeated until the spectrometer tune is satisfactory.<br> | ||
Line 17: | Line 17: | ||
<td>Q<SUP>2</SUP> [GeV<SUP>2</SUP>]</td> | <td>Q<SUP>2</SUP> [GeV<SUP>2</SUP>]</td> | ||
<td>W [GeV]</td> | <td>W [GeV]</td> | ||
− | <td> Rate [Hz] at | + | <td> Rate [Hz] at 20 μA</td> |
− | <td> minutes for | + | <td> minutes for 71k events at 20 μA</td> |
</tr> | </tr> | ||
<tr> | <tr> | ||
<td>2.217</td> | <td>2.217</td> | ||
− | <td>0. | + | <td>0.85</td> |
<td>48.75</td> | <td>48.75</td> | ||
− | <td>1. | + | <td>1.28</td> |
− | <td>1. | + | <td>1.47</td> |
− | <td> | + | <td>119</td> |
− | <td> | + | <td>10</td> |
</tr> | </tr> | ||
</table> | </table> |
Latest revision as of 21:52, 29 September 2016
Procedure without Sieve-slit
- Kinematics need to be updated!
- Time estimate: 1-2 hours
* Take a run at the following kinematics without the sieve-slit collimator. * Start with initial right HRS spectrometer tune with the new quad in place of right Q1. * If you need to increase the spectrometer momentum setting, make sure you cycle Q2 and Q3 as per the cycling procedure. * Electron rates assume 9 out of 11 carbon foils, 5 mSr acceptance, and 8% delta acceptance. * A GMp expert will check the data quality and decide, if the right HRS tune needs to be adjusted to change the size of the acceptance. * This process will be repeated until the spectrometer tune is satisfactory.
Ebeam [GeV] | P0 [GeV/c] | θe [deg] | Q2 [GeV2] | W [GeV] | Rate [Hz] at 20 μA | minutes for 71k events at 20 μA |
2.217 | 0.85 | 48.75 | 1.28 | 1.47 | 119 | 10 |
We want to take runs with the following Q1 currents:
- 0%: 159.75 A
- -6%: 150.165 A
- -5%: 151.7625 A
- -4%: 153.36 A
- -3%: 154.9575 A
- +5%: 167.7375 A