Difference between revisions of "Left RICH"

From Hall A Wiki
Jump to: navigation, search
m (RICH ON-LINE MONITOR SHIFT WORKERS)
 
(8 intermediate revisions by the same user not shown)
Line 1: Line 1:
 
[http://www.iss.infn.it/webg3/cebaf/hadron.html HallA RICH, old page]
 
[http://www.iss.infn.it/webg3/cebaf/hadron.html HallA RICH, old page]
  
== RICH ON-LINE MONITOR SHIFT WORKERS ==
+
[[Shift Checks]] and what to do in case of chamber trips
  
Go on the ''New HallA Tools, "EDM (HLAMain)'' (the old works also fine), click on <u>RICH Detector</u> then check:
+
How to acquire a [[RICH pedestal run]], generate the pedestal files and make them visible to the DAQ, for the next run
* '''Radiator Pressure''' ~ '''10 V''',
+
* '''Power Supplies''' ~ '''3.8 V''' (both polarities: A nearly 3.76 V, B nearly 3.93 V),
+
  
then click on the <u>CAEN</u> buttom;
+
[[RICH Information for experts]]
  
the ''Crate 1'' window pop up, click on <u>Modules</u> and then again on the next window ''Crate #1'' click on <u>Modules</u> of ''Module 2'' (the only active module, then you can check and modify HV).
+
[http://hallaweb.jlab.org/dvcslog/ARICH/ ARICH-elog]
+
Channel 16 and Channel 17 are the only channels connected to the detector.  
+
 
+
They have to be on:
+
* Status "Normal",
+
* HVon have to be "ON"
+
* the voltage "Vmon" has to correspond to the voltage "Vset"
+
  i. nominally between 2200 and 2300 for Channel 16
+
  i. between 200 and 400 on Channel 17.
+
 
+
If not, click on the <u>Module 2 Controls Screen</u> button, on the bottom of the panel, and click on <u>HVon ON</u>, then the HV should raise up very slowly, 5 V/s.
+
 
+
If a trip occurred, the Status of the channel will be ''Trip'', after turning on the HV go back to the first ''Crate 1'' window and click on <u>Clear All Alarm</u>.
+
 
+
Then the status should change into ''Up'' and eventually into ''Normal''.
+
 
+
== Turn HV OFF DURING BEAM STUDIES OR CONFIGURATION CHANGES OF L-HRS. ==
+
 
+
In case of repetitive trips, keep HV OFF also during configuration changes of BigBite.
+
 
+
If the RICH is continuing tripping, keep HV off and call one of the [[RICH experts]],
+
 
+
Do not try to keep HV on more than twice consecutively.
+

Latest revision as of 12:24, 31 January 2009

HallA RICH, old page

Shift Checks and what to do in case of chamber trips

How to acquire a RICH pedestal run, generate the pedestal files and make them visible to the DAQ, for the next run

RICH Information for experts

ARICH-elog