Difference between revisions of "NMR water cell calibration"
From Hall A Wiki
(→Coil Position Survey) |
(→Coil Position Survey) |
||
Line 14: | Line 14: | ||
Cell Diameter = 19.27 +/- 0.05 mm | Cell Diameter = 19.27 +/- 0.05 mm | ||
Distance between 2 pick up coil inner surface : 21.31 +/- 0.14 mm | Distance between 2 pick up coil inner surface : 21.31 +/- 0.14 mm | ||
− | Cell to left Coil Inner Surface = 0.34 mm (could come with | + | Cell to left Coil Inner Surface = 0.34 mm (Preliminary, could come with high Error) |
− | Cell to right Coil Inner Surface = 1.70 mm (could come with | + | Cell to right Coil Inner Surface = 1.70 mm (Preliminary, could come with high Error) |
Revision as of 20:54, 30 September 2008
NMR signal to He3 polarization relation is calibrated by water cell test.
Introduction
Sept 28, 2008 measurement
Coil Position Survey
All error below is 65% CL static error Up Stream: Cell Diameter = 19.27 +/- 0.05 mm Distance between 2 pick up coil inner surface : 21.31 +/- 0.14 mm Cell to left Coil Inner Surface = 0.34 mm (Preliminary, could come with high Error) Cell to right Coil Inner Surface = 1.70 mm (Preliminary, could come with high Error) Down Stream: Cell Diameter = 18.62 +/- 0.03 mm Distance between 2 pick up coil inner surface : 20.87 +/- 0.07 mm