DAC Value 0 1000 2000 3000 4000 5000 6000 7000 8000 9000 10000 11000 12000 13000
 +
Output(V) -4.99 -4.84 -4.69 -4.53 -4.38 -4.23 -4.08 -3.92 -3.77 -3.62 -3.47 -3.32 -3.16 -3.01
 +
 
 +
DAC Value 14000 15000 16000 17000 18000 19000 20000 21000 22000 23000 24000 25000 26000 27000
 +
Output(V) -2.86 -2.70 -2.55 -2.40 -2.25 -2.09 -1.949 -1.797 -1.645 -1.492 -1.340 -1.187 -1.034 -0.882
 +
 
 +
DAC Value 28000 29000 30000 31000 32000 32767 33000 34000 35000 36000 37000 38000 39000 40000
 +
Output(V) -0.729 -0.576 -0.423 -0.270 -0.118 -0.0007 0.034 0.187 0.339 0.492 0.645 0.798 0.951 1.103
 +
 
 +
DAC Value 41000 42000 43000 44000 45000 46000 47000 48000 49000 50000 51000 52000 53000 54000
 +
Output(V) 1.256 1.409 1.562 1.714 1.867 2.01 2.16 2.31 2.47 2.62 2.77 2.92 3.08 3.23
 +
 
 +
DAC Value 55000 56000 57000 58000 59000 60000 61000 62000 63000 64000 65000 65535
 +
Output(V) 3.38 3.53 3.69 3.84 3.99 4.15 4.30 4.45 4.60 4.76 4.91 4.99
 +
 
 +
relationship: Output(v)=0.00015225*DACvalue-4.990
 +
fitting uncertainty: 0.00000003 for slope, 0.0009 for intercept
  
 
==Flexible I/O Setting==
 
==Flexible I/O Setting==

Revision as of 17:38, 2 February 2012

Happex DAC Value(Right Arm)

DAC Value	0	1000	2000	3000	4000	5000	6000	7000	8000	9000	10000	11000	12000	13000
Output(V)	-4.99	-4.84	-4.69	-4.53	-4.38	-4.23	-4.079	-3.927	-3.774	-3.662	-3.47	-3.318	-3.165	-3.013
DAC Value	14000	15000	16000	17000	18000	19000	20000	21000	22000	23000	24000	25000	26000	27000
Output(V)	-2.861	-2.709	-2.556	-2.404	-2.252	-2.099	-1.947	-1.795	-1.642	-1.49	-1.338	-1.185	-1.033	-0.88
DAC Value	28000	29000	30000	31000	32000	32767	33000	34000	35000	36000	37000	38000	39000	40000
Output(V)	-0.728	-0.576	-0.423	-0.271	-0.119	0	0.0337	0.1862	0.3386	0.491	0.643	0.796	0.948	1.101
DAC Value	41000	42000	43000	44000	45000	46000	47000	48000	49000	50000	51000	52000	53000	54000
Output(V)	1.253	1.405	1.558	1.71	1.863	2.015	2.167	2.32	2.472	2.625	2.777	2.929	3.082	3.234
DAC Value	55000	56000	57000	58000	59000	60000	61000	62000	63000	64000	65000	65535
Output(V)	3.387	3.539	3.691	3.844	3.996	4.15	4.3	4.46	4.61	4.76	4.92	4.99
relationship: Output(v)=0.0001524*DACvalue-4.99558

Happex DAC Value(Counting house)

DAC Value	0	1000	2000	3000	4000	5000	6000	7000	8000	9000	10000	11000	12000	13000
Output(V)	-4.99	-4.84	-4.69	-4.53	-4.38	-4.23	-4.08	-3.92	-3.77	-3.62	-3.47	-3.32	-3.16	-3.01
DAC Value	14000	15000	16000	17000	18000	19000	20000	21000	22000	23000	24000	25000	26000	27000
Output(V)	-2.86	-2.70	-2.55	-2.40	-2.25	-2.09	-1.949	-1.797	-1.645	-1.492	-1.340	-1.187	-1.034	-0.882
DAC Value	28000	29000	30000	31000	32000	32767	33000	34000	35000	36000	37000	38000	39000	40000
Output(V)	-0.729	-0.576	-0.423	-0.270	-0.118	-0.0007	0.034	0.187	0.339	0.492	0.645	0.798	0.951	1.103
DAC Value	41000	42000	43000	44000	45000	46000	47000	48000	49000	50000	51000	52000	53000	54000
Output(V)	1.256	1.409	1.562	1.714	1.867	2.01	2.16	2.31	2.47	2.62	2.77	2.92	3.08	3.23
DAC Value	55000	56000	57000	58000	59000	60000	61000	62000	63000	64000	65000	65535
Output(V)	3.38	3.53	3.69	3.84	3.99	4.15	4.30	4.45	4.60	4.76	4.91	4.99
relationship: Output(v)=0.00015225*DACvalue-4.990
fitting uncertainty: 0.00000003 for slope, 0.0009 for intercept

Flexible I/O Setting

To use the trigger interrupt from flexible I/O, the switches must be in the positions:

(D2,D1,D0)    Latching Signal
    0         Input 0
    1         Input 1
    2         Input 2 
    3         Input 3
    4         External Strobe
    5         OR(Input 0,Input 1,Input 2,Input 3)
    6         OR(Input 0,Input 1,Input 2,Input 3,External Strobe)
    7         No latching Signal

Note: open swtich(OFF) is a "1", closed switch(ON) is a "0"

In main board of FlexIO, there is a switch to choose the priority of interrupt.