;This procedure was created for GEN, and must be updated to reflect changes in the target for g2p.
 
;This procedure was created for GEN, and must be updated to reflect changes in the target for g2p.
 +
 +
TO aid in the TE calculation, use the "te_getter.pl" script in the g2p_te folder on poltarg2p2.
 +
 +
Changes for g2p: Set the run valve to a reasonable value, close to the setpoint chosen by the PID. 65% to 90% is a bit too much for this fridge.
  
 
===Assumptions:===
 
===Assumptions:===

Revision as of 17:08, 28 February 2012

This procedure was created for GEN, and must be updated to reflect changes in the target for g2p.

TO aid in the TE calculation, use the "te_getter.pl" script in the g2p_te folder on poltarg2p2.

Changes for g2p: Set the run valve to a reasonable value, close to the setpoint chosen by the PID. 65% to 90% is a bit too much for this fridge.

Assumptions:

  • NMR System is tuned
  • Magnet is at full field, persistent mode, leads are ramped down
  • Nose is full, run valve in manual mode

Prepare Fridge:

  • Stop RB3, RB2, and RB1 (wait 2 mintues between each), if necessary
  • Put Run Valve, EV91120, in Manual Mode
  • Establish a STEADY Nose level somewhere between 65% and 90%
  • Only make slight changes to Run Valve for rest of TE (+/- 0.05 each 10 min)

Take a Baseline:

  • Hit "Unlock Magnet Controls" button
  • Type the full-field current in the box labeled "Setpoint" (eg: 77.085)
  • Type 0.50 into "Setrate" box, if necessary
  • Hit "To Setpoint" Button
  • Wait for leads to reach full current (approx 1 min)
  • Hit "Hold"
  • Check that leads current and magnet current are equal
  • Hit "Heater On" button
  • Wait for the 30 second timeout to expire
  • Type the baseline current into the "Setpoint" box (eg: 74.600)
  • Hit "To Setpoint" to start the magnet sweeping
  • Wait for magnet to reach baseline current (approx 6 min)
  • Hit "Hold" button
  • Put NMR into pause mode if necessary
  • Hit the Baseline button
  • Select "Create New Baseline", dialog box should then disappear
  • Change sweeps to 5000 if necessary
  • Double check that all NMR settings are where you want them
  • Hit "One Point" button to take a single nmr measurement
  • Wait for timer to count down
  • Hit Baseline button
  • Select the baseline you just took from the list of timestamps
  • Document the details of the baseline in the logbook:
    • Date Time, #Sweeps, MagCurrent, Top/Bottom, Gain, RFFreq, RFMod
  • Type full-field current into "Setpoint" box (eg: 77.085)
  • Type 0.50 into "Setrate" box, if necessary
  • Hit "To Setpoint" button
  • Wait for magnet to reach full current (approx 6 min)
  • Hit "Hold" button
  • Hit "Heater Off" button
  • Wait for 30 second timeout to expire
  • Hit "To Zero" to ramp the leads down
  • Hit "Lock Magnet Controls" button

Take TE measurements:

  • Make sure ladder is in desired target position
  • Make sure NMR is on desired channel (AND in agreement with target position!)
  • Set sweeps to 5000, if necessary
  • Hit "Take Data" button
  • Wait for timer to count down
  • Write the following in the logbook for the next 10 measurements:
    • Time, NMR Area, 4He Press, 4He Temp, 3He Press, 3He Temp, Nose Level