1. oven 230, PC 212.4 224.3, TC 34.7 54.3 38.5 | 1. oven 230, PC 212.4 224.3, TC 34.7 54.3 38.5 | ||
spin up 11x120m NMR_Freq_He_20120214_174943 | spin up 11x120m NMR_Freq_He_20120214_174943 | ||
+ | |||
+ | 2. oven 240,PC 225.9,TC 72.0,37.8 | ||
+ | EPR_AFP_20120309_1742 | ||
+ | Polarization 42.07% | ||
+ | Average Frequency Shift 48.28KHz | ||
+ | |||
+ | 2. oven 240,PC 225.9,TC 72.0,37.8 | ||
+ | spin up 12*120m NMR_Freq_He_20120309_181134 | ||
+ | EPR_AFP_20120310_1638 | ||
+ | |||
+ | 3. oven 240,PC 225.9,TC 72.0,37.8 | ||
+ | spin up 12*120m NMR_Freq_He_20120310_164730 | ||
+ | EPR_AFP_20120311_1557 | ||
= test 10 == | = test 10 == |
Latest revision as of 16:05, 11 March 2012
Contents
- 1 Goal
- 2 Previous study
- 3 current run plan
- 4 test 1 (before water calibration)
- 5 water calibration
- 6 test 2 after water calibration
- 7 test 3 after water calibration
- 8 test 4 after water calibration
- 9 test 5 after water calibration and adjust PC pickup coil
- 10 test 6 compare NMR field sweep and Frequency sweep
- 11 test 7, frequency sweep for spin up
- 12 test 8, wait two days to reach max pol
- 13 test 9, after check optics, seal oven etc
- 14 test 10 =
- 15 result
- 16 result outdated
Goal
To optimize temperature for 3He polarization.
Previous study
summer 2010, using cell Boris, done by John Donaghy et al, note available in paper
current run plan
using cell Boris.
use three 30na coherent laser
holding field 25Gs, phi 180
chose 8 pumping cell temp points from 180 to 250C
Then step by step, we do:
1. adjust oven to desired temperature and wait for it to be stable, then turn on laser
(laser will heat up, find the ovan temperature relation with desired pumping cell temp)
2. measure NMR every 2 hours for 12 times (1 day), 3He should reach max pol now.
3. measure EPR once.
4. measure AFP loss. turn off laser, measure NMR every 2 minutes for 5 times.
5. measure depolarization. keep laser off, measure NMR every 1 hour for 10 times.
test 1 (before water calibration)
- 1. temperature: oven 210, pumping cell 217, target cell 79.9 (217 1st round, 217 2nd round, 217 3rd round, 217 4th round,217 5th round)
- 2. temperature: oven 230, pumping cell 233, target cell 86.7 (233 1st round,233 2nd round,233 3rd round)
- 3. temperature: oven 240, pumping cell 233, target cell 86.7 (NMR_He_20110129_010725 20x120m, spin up, oven temp stable at 233, can't go higher due to air leak at ovan bottom)
- 4. temperature: oven 220, pumping cell 223, target cell 83.7 (223 1st round,223 2nd round)
- 5. temperature: oven 225, pumping cell 228, target cell 84.8 (228 1st round,228 2nd round)
- 6. temperature: oven 215, pumping cell 218.6, target cell 82.5 (218 1st round)
- 7. temperature: oven 190, pumping cell 195, target cell 74.1 (195 1st round,195 2nd round)
- 8. temperature: oven 190, pumping cell 196.2, target cell 75.2 (196 1st round)
- 9. temperature: oven 170, pumping cell 176.8, target cell 67.5 (176 1st round)
water calibration
test 2 after water calibration
test 3 after water calibration
- increase temp right after EPR without destroying pol
1. 170 oven, 186.5 PC, 27.7-27.8 38.1-39.3 30.6-30.8 TC
spin up 11x120m NMR_He_20111119_162441 onetime NMR_He_20111125_182119 (after 5 days with laser on) EPR EPR_AFP_20111125_1835
2. 190 oven, 204.3 PC, 28.4-28.6 40.5-41.2 31.4-31.6 TC
spin up 10x120m NMR_He_20111125_184220 onetime NMR_He_20111126_214648 EPR EPR_AFP_20111126_2215
3. 210 oven, 222.4 PC, 29.6 44.9-45.2 32.6-32.7 TC
spin up 8x180m NMR_He_20111126_222112 onetime NMR_He_20111127_204311 EPR EPR_AFP_20111127_2058
4. oven 215, PC 226.8, 30.0 45.6 32.7 TC
spin up 10x120m NMR_He_20111127_210342 onetime NMR_He_20111128_200905 EPR EPR_AFP_20111128_2030
5. oven 220, 231.3 PC, 30.2 45.1 34.4 TC
spin up setting is wrong, no data onetime NMR_He_20111129_135749 onetime NMR_He_20111129_172355 EPR EPR_AFP_20111129_1749
5. oven 225
spin up NMR_He_20111129_175728
test 4 after water calibration
1. oven 220, PC 231.9, TC 29.9 45.2 33.6
spin up 11*120m NMR_He_20111130_200754 EPR EPR_AFP_20111201_1622
NMR 5*0.4m NMR_He_20111201_163939 laser on NMR 5*2m NMR_He_20111201_164607 laser on NMR 5*0.4m NMR_He_20111201_165843 laser off NMR 5*2m NMR_He_20111201_170204 laser off
2. oven 225, PC 236.1, TC 30.1 45.7 34.3
EPR EPR_AFP_20111202_1606 NMR 5*0.4m NMR_He_20111202_161802 laser on NMR 5*2m NMR_He_20111202_162047 laser on NMR 5*0.4m NMR_He_20111202_163137 laser off NMR 5*2m NMR_He_20111202_163820 laser off
New Test: oven 225, PC 236.0, TC 30.0~30.1,45.9~46.8,33.8~34.3 spin up 12*120m NMR_He_20111202_173326 EPR EPR_AFP_20111203_1620
laser off NMR_field 5*0.4m NMR_He_20111203_163043 NMR_field 5*1m NMR_He_20111203_163352 NMR_field 5*2m NMR_He_20111203_163826 NMR_freq 5*0.4m NMR_Freq_He_20111203_164910 NMR_freq 5*1m NMR_Freq_He_20111203_165115 NMR_freq 5*2m NMR_Freq_He_20111203_165548
laser on NMR_field 5*0.4m NMR_He_20111203_170646 NMR_field 5*1m NMR_He_20111203_170929 NMR_field 5*2m NMR_He_20111203_171420 NMR_field 5*0.4m NMR_He_20111203_172304 NMR_freq 5*0.4m NMR_Freq_He_20111203_172848 NMR_freq 5*1m NMR_Freq_He_20111203_173047
3. oven 230, PC 240.4, TC 30.8-30.99 46.8-47.0 34.2-34.4
spin up 11x120m NMR_He_20111203_173536 (polarization was not destroyed before spin up) EPR EPR_AFP_20111204_1934
laser on NMR_field 5*0.4m NMR_He_20111204_185005
laser off and wait for 15m, temp of PC stay at 203 NMR_field 5x1m NMR_He_20111204_200414 NMR_freq 5x1m NMR_Freq_He_20111204_201331
test 5 after water calibration and adjust PC pickup coil
1. oven 230 T_pumping=239.3C, T_target=32.1C, 47.9C, 35.7C
spin up 12x120m NMR_He_20111206_164213 (PC amp 48.9699 amp_error 0.246622 tau 17.6184 tau_error 0.166278) (TC amp 38.8611 amp_error 0.122741 tau 16.678 tau_error 0.0980802) EPR EPR_AFP_20111207_1550
AFP loss NMR_He_20111207_161123 5x1min, laser on, good data (PC loss 0.01189 loss 0.0104945 loss 0.00876603 loss 0.011773 loss_ave 0.0107309) (TC loss 0.0107156 loss 0.00943059 loss 0.00876428 loss 0.0134604 loss_ave 0.0105927) NMR_He_20111207_161851 5x2min, laser on, need to be checked NMR_He_20111207_164230 5x40s, laser on NMR_He_20111207_164230 5x1min, laser on
T_pumping=203C, T_target=31.9C, 48.4C, 34.7C laser off NMR_He_20111207_170036 5x1min, laser off, good data NMR_He_20111207_170818 5x2min, laser off
2. oven 235, cant reach 235 even after adjusting air flow
spin up 11x120m NMR_He_20111207_173118
3. oven 225, PC 234.8, TC 31.4 47.0 34.6
spin up 12x120m NMR_He_20111208_150200 (PC amp 38.8015 amp_error 0.21706 tau 17.7425 tau_error 0.1844) (TC amp 32.1239 amp_error 0.1227 tau 17.5481 tau_error 0.124282) EPR EPR_AFP_20111209_1635 AFP loss NMR_He_20111209_164421 5x1min, laser on NMR_He_20111209_165305 5x1min, laser on NMR_He_20111209_170044 5x1min, laser on NMR_He_20111209_171519 5x1min, laser on good one (PC loss 0.00253009 loss 0.001489 loss 0.00658421 loss 0.000295381 loss_ave 0.00272467) (TC loss 0.00246562 loss 0.00191102 loss 0.00229629 loss 0.0044294 loss_ave 0.00277558)
4. oven 220, PC 230.1, TC 30.7 45.9 35.0
spin up 12x120m NMR_He_20111209_223141 (PC amp 35.7233 amp_error 0.265447 tau 19.0275 tau_error 0.254762) (TC amp 29.6341 amp_error 0.137933 tau 19.2899 tau_error 0.159346) EPR EPR_AFP_20111210_2137 AFP loss 5x1m laser on NMR_He_20111210_214037 NMR_He_20111210_214622 5x1min, laser on NMR_He_20111210_215200 5x1min, laser on good one (PC loss 0.00304883 loss 0.00658595 loss -0.00224954 loss 0.00703551 loss_ave 0.00360519) (TC loss 0.00152955 loss 0.00417518 loss 0.00490693 loss 0.00108878 loss_ave 0.00292511)
5. oven 210,PC 220.8 TC 30.0 44.9 33.0
spin up 11x120m NMR_He_20111210_215903 (PC amp 28.1674 amp_error 0.0577626 tau 30 tau_error 5.15775) (TC amp 22.7238 amp_error 0.253823 tau 22.5265 tau_error 2.3058) EPR EPR_AFP_20111211_1841 AFP loss 5x1m laser on NMR_He_20111211_184542 NMR_He_20111211_185142 NMR_He_20111211_185947 good one (PC loss 0.00361561 loss 0.00329046 loss 0.00382612 loss 9.61192e-05 loss_ave 0.00270708) (TC loss -0.000158035 loss 0.00543516 loss 0.00263695 loss 0.00288104 loss_ave 0.00269878)
6. oven 230, PC 239.3, TC 31.8 48.0 35.3
spin up 11x120m NMR_He_20111211_191237 onetime NMR_He_20111212_153333 onetime NMR_Freq_He_20111212_153603 (RF 3.2V) EPR EPR_AFP_20111212_1613 onetime NMR_He_20111212_161958 onetime NMR_Freq_He_20111212_163526 (RF 1.5V) AFP 5x1m laser on NMR_He_20111212_164824
7. oven 230
spin up 11x120m NMR_He_20111212_181326
- note:
- The sweep interval counting starts from beginning of every sweep, so the sweep time is included.
- one field sweep takes about 30s.
- one frequency sweep takes about 10s.
- after laser off, temp reach stables in about 15m
test 6 compare NMR field sweep and Frequency sweep
oven 230
NMR_Freq_He_20111214_164342 RF 1.5V amp_PC 19.0443 0.106143 NMR_He_20111214_164514 RF 3.2V amp_PC 30.8823 0.0384327 ratio 1.62
NMR_He_20111215_003810: 13.6374 0.0341254 NMR_Freq_He_20111215_004032: 8.15775 0.468220 ratio1.6717
NMR_He_20111215_111133 21.42 0.04848 NMR_Freq_He_20111215_111253 12.98 0.07145 ratio1.6502
test 7, frequency sweep for spin up
1. oven 230
spinup NMR_Freq_He_20111215_235120
2. oven 225, PC 233.9, TC 31.5 47.9 34.9
spinup 12x120m NMR_Freq_He_20111216_175821 RF1.5V onetime NMR_Freq_He_20111217_190903 RF1.5V onetime NMR_He_20111217_191208 RF3.2V EPR EPR_AFP_20111217_1921 RF1.5V AFP loss laserOn 5x1m NMR_Freq_He_20111217_192629 RF1.5V NMR_Freq_He_20111217_193223 RF1.5V NMR_Freq_He_20111217_193714 RF1.5V good
3. oven 220, PC 229.4, TC 31.4 47.5 34.9
spinup 11x120m NMR_Freq_He_20111217_194817 RF1.5V onetime NMR_Freq_He_20111218_173756 RF1.5V onetime NMR_He_20111218_173844 RF3.2V EPR EPR_AFP_20111218_1851 bad, freqency fluctuate a lot, can't lock AFP loss laserOn 5x1m NMR_Freq_He_20111218_181336 RF1.5V NMR_Freq_He_20111218_181814 RF1.5V NMR_Freq_He_20111218_182337 RF1.5V NMR_Freq_He_20111218_183245 RF1.5V good NMR_Freq_He_20111218_183245 RF1.5V
4. oven 210
spinup 11x120m NMR_Freq_He_20111218_185511 RF1.5V onetime NMR_Freq_He_20111219_173126 RF1.5V onetime NMR_He_20111219_173238 EPR AFP loss laser on 5*1m NMR_Freq_He_20111219_173658 RF1.5V NMR_Freq_He_20111219_174325
test 8, wait two days to reach max pol
1. oven 220, PC 228.7 TC 30.6 47.1 33.0
spinup 6x500m NMR_Freq_He_20120110_175457 RF1.5V onetime NMR_He_20120112_154548 RF3.2V EPR EPR_AFP_20120112_1638
test 9, after check optics, seal oven etc
1. oven 230, PC 212.4 224.3, TC 34.7 54.3 38.5
spin up 11x120m NMR_Freq_He_20120214_174943
2. oven 240,PC 225.9,TC 72.0,37.8
EPR_AFP_20120309_1742 Polarization 42.07% Average Frequency Shift 48.28KHz
2. oven 240,PC 225.9,TC 72.0,37.8
spin up 12*120m NMR_Freq_He_20120309_181134 EPR_AFP_20120310_1638
3. oven 240,PC 225.9,TC 72.0,37.8
spin up 12*120m NMR_Freq_He_20120310_164730 EPR_AFP_20120311_1557
test 10 =
result
Nov15,2011Media:Nov15.pdf
PPT ON Aug.31,2011 including the fitted tao for target cell and pumping cell:Media:Aug31.pdf
result outdated
pumpint chamber target chamber test max. pol tao max. pol tao 217 spin up 22.9028 0.1705 3.88819 0.1914 spin down 0.1025 0.09846 233 spin up 26.8936 0.2733 4.52142 0.2554 spin down 0.147 0.147 223 spin down 27.7217 0.1169 4.83401 0.1081