+ | <pre> | ||
+ | All error below is 65% CL static error | ||
+ | |||
+ | Up Stream: | ||
+ | Cell Diameter = 19.27 +/- 0.05 mm | ||
+ | Distance between 2 pick up coil inner surface : 21.31 +/- 0.14 mm | ||
+ | |||
+ | Down Stream: | ||
+ | Cell Diameter = 18.62 +/- 0.03 mm | ||
+ | Distance between 2 pick up coil inner surface : 20.87 +/- 0.07 mm | ||
+ | </pre> |
Revision as of 15:50, 29 September 2008
NMR signal to He3 polarization relation is calibrated by water cell test.
Introduction
Sept 28, 2008 measurement
Coil Position Survey
All error below is 65% CL static error Up Stream: Cell Diameter = 19.27 +/- 0.05 mm Distance between 2 pick up coil inner surface : 21.31 +/- 0.14 mm Down Stream: Cell Diameter = 18.62 +/- 0.03 mm Distance between 2 pick up coil inner surface : 20.87 +/- 0.07 mm