Difference between revisions of "Beam Line"

From Hall A Wiki
Jump to: navigation, search
(Beam position monitors (BPM) and calibration procedure)
(Beam position monitors (BPM) and calibration procedure)
Line 26: Line 26:
 
     - from there pull down the "Expert Screens" and open "Gain Control"
 
     - from there pull down the "Expert Screens" and open "Gain Control"
 
     - for IOCSE10, use the pull down window and change from "auto gain" to "forced gain".  
 
     - for IOCSE10, use the pull down window and change from "auto gain" to "forced gain".  
     - then change the Forced Gain values x and y to zero.
+
     - then change the forced gain values x and y to zero.
  
 
2.  CODA has to run in pedestal mode:
 
2.  CODA has to run in pedestal mode:
Line 33: Line 33:
 
     - select "PEDRUN"  - Press RunTypes button in RunControl
 
     - select "PEDRUN"  - Press RunTypes button in RunControl
  
3.  Run [http://hallaweb.jlab.org/experiment/E97-110/get_bpm_pedestals.C get_bpm_pedestals.C]
+
3.  Have MCC change the BPM gain back to "auto gain".
 +
 
 +
4.  Run [http://hallaweb.jlab.org/experiment/E97-110/get_bpm_pedestals.C get_bpm_pedestals.C]
 
in the root analyzer:  
 
in the root analyzer:  
 
     - you might have to modify some details  
 
     - you might have to modify some details  

Revision as of 17:22, 2 October 2008

E06-010 Instrumentation Requirement List: Jian-Ping Chen's memo to Kees De Jager, 08/07/08

.doc file and htm file

ARC energy measurement

Compton Polarimeter

HARP and HARP scan procedures

Please follow the directions at the Harp Scan Page [Harp Scan page]

Beam raster

Beam charge monitors (BCM) and calibration procedure

Beam position monitors (BPM) and calibration procedure

How to perform a BPM pedestal run:

1. MCC hast to follow the following procedure:

    - open BPM window "BPM Diagnostics - SEE"
    - from there pull down the "Expert Screens" and open "Gain Control"
    - for IOCSE10, use the pull down window and change from "auto gain" to "forced gain". 
    - then change the forced gain values x and y to zero.

2. CODA has to run in pedestal mode:

    - if you were running before you first have to press Abort
    - choose the appropriate configuration in runcontrol
    - select "PEDRUN"  - Press RunTypes button in RunControl

3. Have MCC change the BPM gain back to "auto gain".

4. Run get_bpm_pedestals.C in the root analyzer:

    - you might have to modify some details 
    - the code should print out the new pedestals at the end 
    - load the database into an editor and change them (2nd line of the [BPMx] blocks)