This page summarize the calibration data for quarter wave plate holders, which control the pumping laser polarization of Hall A He3 Polarized Target.

Connection Map

Pumping Direction Laser Component Motor ID Laser Platform Patch Panel 50Ft Ext. Cable ID Controller ID
Longitudinal S Wave 1 (tested & taken out) 1 1 None
Longitudinal P Wave 2 (tested & taken out) 2 2 None
Transverse S Wave 3 3 3 0338
Transverse P Wave 4 4 4 1414
Vertical S Wave 5 5 5 1426
Vertical P Wave 6 6 6 1423

Current Setting

Motor ID Controller ID Description Alignment Angle/Degree Auto calibration Error/Degree Spin Up Read Back/Degree Spin Down Read Back/Degree
1  ?  ? 8.4 0.15 8.25 -81.75
2  ?  ? 8.5 0.40 8.10 -81.90
3  ?  ? 7.5 -1.50 9.00 -81.00
4  ?  ? 8.8 0.00 8.80 -81.20
5  ?  ? 10.8 0.40 10.40 -79.60
6  ?  ? 9.8 1.00 8.80 -81.20

  • Motor ID : The sticker attached to the motor
  • Controller ID : ID number of control connected to the motor
  • Alignment Angle : When the mark of Q-wave plate is at this angle, all plates will convert a vertical polarized laser to a circular polarized one. Besides, the polarization direction is the same.
  • Auto calibration error : is the marker angle on the plate holder, while the holder is shown to be rotated to 0 degree on the computer. The reason why it is not 0, is due to non-perfect precision of automatic calibration.
  • Position Up/Down Read Back: The angle setting on the computer for spin state Up/Down.

Old Settings