Single FAP Lasers

DUO FAP Lasers

COMET Lasers

Power Test Result of COMET Lasers

Test Condition

  • Room Temperature: 22.9 C
  • Diode Temperature: 20.0 C

W&M COMET Laser

Diode Current (I) Driving Voltage (V) Cooling Current (I) Laser Power (W) Central Wavelength (nm) Wavelength FWHM (nm)
10 1.67 1.56 2.67 794.84 0.55
15 1.74 1.99 7.18 794.90 0.53
20 1.80 2.45 11.9 794.98 0.52
25 1.87 2.96 16.5 795.03 0.50
30 1.93 3.56 21.1 795.09 0.52
35 1.99 4.29 25.6 795.15 0.56
40 2.05 5.18 30.3 795.22 0.60

Wm comet.png

Rutgers COMET Laser

Diode Current (I) Driving Voltage (V) Cooling Current (I) Laser Power (W) Central Wavelength (nm) Wavelength FWHM (nm)
10 1.66 1.71 1.34 794.79 0.58
15 1.72 2.18 5.82 794.83 0.50
20 1.78 2.57 10.5 794.88 0.51
25 1.84 2.77 15.1 794.91 0.53
30 1.90 3.31 19.7 794.94 0.57
35 1.95 4.18 24.3 795.01 0.57
40 2.01 4.95 28.6 795.06 0.62
44 2.05 5.53 32.1 795.11 0.64

Rutgers comet.png

Laser Alignment