<ol> | <ol> | ||
<li>Start '''dbedit''' on '''adaq1''', select "'''mollerpol_l'''" database and set in "'''beam_pol'''" configuration table in row "'''code'''" | <li>Start '''dbedit''' on '''adaq1''', select "'''mollerpol_l'''" database and set in "'''beam_pol'''" configuration table in row "'''code'''" | ||
Line 16: | Line 17: | ||
</pre> | </pre> | ||
</li> | </li> | ||
+ | == Executable Kumac's == | ||
+ | === Run.kumac === |
Revision as of 15:56, 13 August 2018
Making Polarization measurements
Overview
- Start dbedit on adaq1, select "mollerpol_l" database and set in "beam_pol" configuration table in row "code" parameter "adcevents" from 9 to 2: " ... adcevents=2"
- Make at least two CODA runs in order to use both coil polarities.
- Data analysis:
- old DAQ: adaq1:~/paw/moller/
> paw paw> exec run run=15625 paw> exec lg_spectra run=15625 paw> exec time run=15625 paw> lcd mag_optim paw> exec anpow_set e0=7.37 (E0 - beam energy in GeV) paw> exec asym11 anpow=0.7735 gate=0.033 run=15625 (gate=0.033 for 30Hz helicity frequency and gate=0.0008 for 1kHz)