Difference between revisions of "LHRS for GMn"

From Hall A Wiki
Jump to: navigation, search
(Nominal Settings)
Line 6: Line 6:
 
* The LHRS dipole NMR doesn't lock. Need to set manually to match the field value (iterative procedure, see the instruction below)
 
* The LHRS dipole NMR doesn't lock. Need to set manually to match the field value (iterative procedure, see the instruction below)
 
* Below are set points. Readbacks will be a bit lower.
 
* Below are set points. Readbacks will be a bit lower.
 +
 +
= Setting the LHRS Magnets =
 +
 +
If one can not reset the magnet remotely: '''Do NOT go into the hall to reset by yourself. Contact RC and then tech on-call/magnet expert.'''
 +
Regulators for Q1, Q2, Q3 are always OFF. Q1, Q2, Q3 magnets will be set by current.
  
 
= LHRS Dipole =  
 
= LHRS Dipole =  
Line 12: Line 17:
  
 
= Checking and Controlling Drifts =
 
= Checking and Controlling Drifts =
 +
 +
= Legacy Pages =
 +
 +
* [[LHRS]]

Revision as of 11:58, 9 June 2021

Back to How-To

Nominal Settings

  • When starting a shift, it is prudent to verify that the magnets are set correctly. If it's not right, consult the RC. Currents should be within about 0.5 A of the values. Also, note that "readback" values are slightly different from "set point" values.
  • The LHRS dipole NMR doesn't lock. Need to set manually to match the field value (iterative procedure, see the instruction below)
  • Below are set points. Readbacks will be a bit lower.

Setting the LHRS Magnets

If one can not reset the magnet remotely: Do NOT go into the hall to reset by yourself. Contact RC and then tech on-call/magnet expert. Regulators for Q1, Q2, Q3 are always OFF. Q1, Q2, Q3 magnets will be set by current.

LHRS Dipole

Monitoring

Checking and Controlling Drifts