Difference between revisions of "HV and Threshold calibration"

From Hall A Wiki
Jump to: navigation, search
Line 4: Line 4:
  
 
* calibrate all neutron bars to equal gains.
 
* calibrate all neutron bars to equal gains.
* set threshold voltage as low as posibble (50 mV = channel 300), to reduce current on PMT's.
+
* set gains as low as posibble, to reduce current on PMT's.
 
* set threshold energy as low as possible to increase efficiency.
 
* set threshold energy as low as possible to increase efficiency.
 
* set threshold energy low enough to remove minimum ionizing electrons.
 
* set threshold energy low enough to remove minimum ionizing electrons.
  
   optimum was chosen at threshold 5 MeV - still need to verify this.
+
   optimum was chosen at threshold = 5 MeV = 300 channels - need to re-check.
  
 +
  this is equivalent to 1 MeV = 60 channels
 +
 
  
 
Calibrations steps -  
 
Calibrations steps -  
Line 19: Line 21:
 
adjust HV, so that the muon peak will be at-
 
adjust HV, so that the muon peak will be at-
  
10.0 cm bar -> E=20 MeV ->  channel=1500
+
10.0 cm bar -> E=20 MeV ->  channel=1200
  
12.5 cm bar -> E=25 MeV ->  channel=1875
+
12.5 cm bar -> E=25 MeV ->  channel=1500
  
15.0 cm bar -> E=30 MeV ->  channel=2250
+
15.0 cm bar -> E=30 MeV ->  channel=1800
  
25.0 cm bar -> E=50 MeV ->  channel=3750
+
25.0 cm bar -> E=50 MeV ->  channel=3000
  
  
 
* Beam calibration using the H(e,e'p) reaction.
 
* Beam calibration using the H(e,e'p) reaction.
  
Use the H(e,e'p) reaction to create 1 GeV/c protons which will have energy loss of 32 MeV for a 10cm wide bar, calibrate the gain cofficients, so these events will be at channel=2000
+
Use the H(e,e'p) reaction to create 1 GeV/c protons which will have energy loss of 32 MeV for a 10cm wide bar, calibrate the gain cofficients, so these events will be at channel=1920

Revision as of 08:46, 13 April 2010

Procedure for the neutron array HV and threshold calibration

Main considerations -

  • calibrate all neutron bars to equal gains.
  • set gains as low as posibble, to reduce current on PMT's.
  • set threshold energy as low as possible to increase efficiency.
  • set threshold energy low enough to remove minimum ionizing electrons.
 optimum was chosen at threshold = 5 MeV = 300 channels - need to re-check.
 this is equivalent to 1 MeV = 60 channels
 

Calibrations steps -

  • Test-area calibration using cosmic muons

Use top-buttom bar trigger, to obtain vertical minimum ionizing muons with known energy deposit.

adjust HV, so that the muon peak will be at-

10.0 cm bar -> E=20 MeV -> channel=1200

12.5 cm bar -> E=25 MeV -> channel=1500

15.0 cm bar -> E=30 MeV -> channel=1800

25.0 cm bar -> E=50 MeV -> channel=3000


  • Beam calibration using the H(e,e'p) reaction.

Use the H(e,e'p) reaction to create 1 GeV/c protons which will have energy loss of 32 MeV for a 10cm wide bar, calibrate the gain cofficients, so these events will be at channel=1920