Difference between revisions of "Data Log"

From Hall A Wiki
Jump to: navigation, search
Line 278: Line 278:
  
 
|-
 
|-
| 130||S||Center||3700||||||
+
| 130||S||Center||||3700    ||||
  
 
|-
 
|-
Line 290: Line 290:
  
 
|-
 
|-
|    134||  S    ||  ""    || 3750   ||       ||        || Y reading Bad
+
|    134||  S    ||  ""    ||    || 3750      ||        || Y reading Bad
  
 
|-
 
|-

Revision as of 09:39, 12 July 2010

RUN # Source/Cosmic/Pedestal Source Position Time HV # of events notes
23 S lo x lo y n/a n/a
24 S mid x lo y n/a n/a
25 S hi x lo y n/a n/a
26 S hi x mid y n/a n/a
27 S hi x hi y n/a n/a
28 S mid x mid y n/a n/a
29 C n/a 21 hrs n/a 72383 ROC not responding
35 C n/a n/a 3500
37 C n/a n/a 3550
38 C n/a n/a 3600
39 C n/a n/a 3650
40 C n/a n/a 3700
41 C n/a n/a 3750
42 C n/a n/a 3800
43 C n/a n/a 3600
44 P n/a n/a off
45 P n/a n/a off Ped reconfigured
46 S n/a n/a 3600 7863
47 C n/a n/a 3600
51 C n/a 24.5 hrs 3600 20,000
52 P n/a n/a off 11,000
53 failed
54 stopped early
55 stopped early
56 failed
57 3600
58 C n/a 5.2 hrs 3600 Client unresponsive
59 S n/a n/a 3600
60 S n/a n/a 3600
74 S n/a n/a 3700
75 P n/a n/a off GEM # 3
76 failed
78 P n/a n/a off new read out card
80 C n/a n/a 3700
81 S center n/a 3700
82 S center n/a 3700 rate was abnormally high. Reset veto
83 P n/a n/a off new LV distribution board
84 S center n/a 3700
85 Test with VME Crate Transfer - GEM 2
86 ""
87 ""
88 ""
89 ""
90 ""
91 ""
92 ""
93 ""
94 ""
95 ""
96 P
97 S No HV - bad
98 No Data recorded - Cause was found due to cards with same address
99 P
100 S Center 3700
101 S Low X, Low Y 3700 Bad reading in Y
102 S High X, High Y 3700 No Data recorded - ROC Froze
103 S "" "" No Data recorded - ROC Froze
104 S "" "" Reset ROC
105 S High X, Low Y 3700 Bad reading in Y
106 S High X, Mid Y "" Y Reading Fine
107 S High X, Low-Mid y "" Y Reading Bad
108 P Changed Y Card - Card was not connected
109 P
110 S High X, Low-Mid Y Y Reading Bad
111 ABORTED
112 S Center Old readout Card - It is bad.
113 S Center ROC FROZE
114 Checked CH1 readout - Nothing
115 Checked CH0
116 S Mid X, High Y Y reading Fine
117 P GEM 3<----- PLOTS
118 S Center 3700
119 S Low X, Low Y "" Y reading bad
120 S High X, High Y "" Y Reading fine
121 P Redid Ped Run
122 S High X, Low Y 3700 Y reading Bad
123 P Switched Boards on GEM
124 S High X, Low Y 3700 Y reading Bad
125 Changed Amplifier Output to Slot 2 and messed with DC Voltage
126 P
127 S High X, Low Y 3700 Y reading Bad
128 P GEM 4 <----- PLOTS
129 C 3600 BAD
130 S Center 3700
131 S Low X, Low Y 3700 Y Reading Bad
132 S High X, Mid Y 3700 Y reading good
133 S High X, Mid-Low Y "" Y reading Bad
134 S "" 3750 Y reading Bad
135 S High X, High Y 3700 Y reading good
136 S High X, High Y 3700 Y reading Good, Higher up on Y