Difference between revisions of "Data Log"

From Hall A Wiki
Jump to: navigation, search
Line 333: Line 333:
 
|-
 
|-
 
| 172    ||  S    || Low X, Low Y    ||      ||  3700    ||        ||
 
| 172    ||  S    || Low X, Low Y    ||      ||  3700    ||        ||
 +
 +
|-
 +
|  173  ||      ||      ||      ||      ||        ||aborted
 +
 +
|-
 +
|  174  ||sS    || Low X, Low Y    ||      ||  3700    ||        ||changed to channel 2
 +
 +
|-
 +
|  175  ||  P    ||      ||      ||      ||        ||replaced Y card and conector
 +
 +
|-
 +
|  176  ||  P    ||      ||      ||      ||        ||changed CRAM card
 +
 +
|-
 +
|  177  ||  S  || Mid X, Mid Y    ||      ||  3700    ||        ||
 +
 +
|-
 +
|  178  ||    P  ||      ||      ||      ||        ||GEM 2
 +
 +
|-
 +
|  179  ||    C  ||      ||      ||    3600  ||        ||
 +
 +
|-
 +
|  180  ||      ||      ||      ||      ||        ||changed T.gem.X1.detmap & T.gem.Y1.detmap ->slot_lo to 1
 +
 +
|-
 +
|  181  ||  S    ||      ||      ||      ||        ||changed T.gem.X1.detmap & T.gem.Y1.detmap ->slot_lo to 2 (changed back)
 +
 +
|-
 +
| 182    ||  S    ||      ||      ||      ||        ||
 +
 +
|-
 +
|  183  ||      ||      ||      ||      ||        ||changed T1 to 50 in CRL code for 551b
 +
 +
|-
 +
|  184  ||  P  ||      ||      ||      ||        ||(kaptons unpluged)
 +
 +
|-
 +
|    185 || P    ||      ||      ||      ||        ||
 +
 +
|-
 +
|  186  ||  S  ||      ||      ||    3700  ||        ||Changed T1 from 50 to 40 in CRL code
 +
 +
|-
 +
|  187  ||  S  ||      ||      ||  3700    ||        ||CRL code still changed to 40. Data is going to neg(-) input in 550
 +
 +
|-
 +
|  188  || P    ||      ||      ||      ||        ||Changed polarity of trigger. Hold -> no difference
 +
 +
|-
 +
|  189  ||    S  ||      ||      ||    3700  ||        ||
 +
 +
|-
 +
| 190    ||  S  ||      ||      ||    3700  ||        ||Changed CRL T1->70
 +
 +
|-
 +
|  191  ||      ||      ||      ||    3700  ||        ||Changed CRL T1->60
 +
 +
|-
 +
|  192  ||      ||      ||      ||      ||        ||
 +
 +
|-
 +
|  193  ||      ||      ||      ||      ||        ||Changed CRL T1-> 50. aborted
 +
 +
|-
 +
|    194 ||      ||      ||      ||      ||        ||gate generator to make trigger hold. T1->50
 +
 +
|-
 +
| 195    ||  S    ||      ||      ||    3700  ||        ||T1 50 -> 30
 +
 +
|-
 +
|  196  ||  S  ||      ||      ||    3700  ||        ||T1 30 -> 20
 +
 +
|-
 +
|  197  ||      ||      ||      ||  3750    ||        ||T1 20 -> 70
 +
 +
|-
 +
|  198  ||      ||      ||      ||      ||        ||Clear coming from gate generator
  
 
|-
 
|-
Line 348: Line 426:
 
|-
 
|-
 
|    ||      ||      ||      ||      ||        ||
 
|    ||      ||      ||      ||      ||        ||
 +
 +
|-
 +
|    ||      ||      ||      ||      ||        ||
 +
 +
|-
 +
|    ||      ||      ||      ||      ||        ||
 +
 +
|-
 +
|    ||      ||      ||      ||      ||        ||
 +
 +
|-
 +
|    ||      ||      ||      ||      ||        ||
 +
 +
|-
 +
|    ||      ||      ||      ||      ||        ||
 +
 +
|-
 +
|    ||      ||      ||      ||      ||        ||
 +
 +
|-
 +
|    ||      ||      ||      ||      ||        ||
 +
 +
|-
 +
|    ||      ||      ||      ||      ||        ||
 +
 +
|-
 +
|    ||      ||      ||      ||      ||        ||
 +
 +
|-
 +
|    ||      ||      ||      ||      ||        ||
 +
 +
|-
 +
|    ||      ||      ||      ||      ||        ||
 +
 +
|-
 +
|    ||      ||      ||      ||      ||        ||
 +
 +
|-
 +
|    ||      ||      ||      ||      ||        ||
 +
 +
 +
 +
 +
 +
 +
 +
 +
 +
  
 
|}
 
|}

Revision as of 16:27, 9 September 2010

RUN # Source/Cosmic/Pedestal Source Position Time HV # of events notes
23 S lo x lo y n/a n/a GEM 1 <-----PLOTS
24 S mid x lo y n/a n/a
25 S hi x lo y n/a n/a
26 S hi x mid y n/a n/a
27 S hi x hi y n/a n/a
28 S mid x mid y n/a n/a
29 C n/a 21 hrs n/a 72383 ROC not responding
35 C n/a n/a 3500
37 C n/a n/a 3550
38 C n/a n/a 3600
39 C n/a n/a 3650
40 C n/a n/a 3700
41 C n/a n/a 3750
42 C n/a n/a 3800
43 C n/a n/a 3600
44 P n/a n/a off
45 P n/a n/a off Ped reconfigured
46 S n/a n/a 3600 7863
47 C n/a n/a 3600
51 C n/a 24.5 hrs 3600 20,000
52 P n/a n/a off 11,000
53 failed
54 stopped early
55 stopped early
56 failed
57 3600
58 C n/a 5.2 hrs 3600 Client unresponsive
59 S n/a n/a 3600
60 S n/a n/a 3600
74 S n/a n/a 3700
75 P n/a n/a off GEM # 3
76 failed
78 P n/a n/a off new read out card
80 C n/a n/a 3700
81 S center n/a 3700
82 S center n/a 3700 rate was abnormally high. Reset veto
83 P n/a n/a off new LV distribution board
84 S center n/a 3700
85 Test with VME Crate Transfer - GEM 2
86 ""
87 ""
88 ""
89 ""
90 ""
91 ""
92 ""
93 ""
94 ""
95 ""
96 P
97 S No HV - bad
98 No Data recorded - Cause was found due to cards with same address
99 P
100 S Center 3700 GEM 2 <----- PLOTS
101 S Low X, Low Y 3700 Bad reading in Y
102 S High X, High Y 3700 No Data recorded - ROC Froze
103 S "" "" No Data recorded - ROC Froze
104 S "" "" Reset ROC
105 S High X, Low Y 3700 Bad reading in Y
106 S High X, Mid Y "" Y Reading Fine
107 S High X, Low-Mid y "" Y Reading Bad
108 P Changed Y Card - Card was not connected
109 P
110 S High X, Low-Mid Y Y Reading Bad
111 ABORTED
112 S Center Old readout Card - It is bad.
113 S Center ROC FROZE
114 Checked CH1 readout - Nothing
115 Checked CH0
116 S Mid X, High Y Y reading Fine
117 P GEM 3<----- PLOTS
118 S Center 3700
119 S Low X, Low Y "" Y reading bad
120 S High X, High Y "" Y Reading fine
121 P Redid Ped Run
122 S High X, Low Y 3700 Y reading Bad
123 P Switched Boards on GEM
124 S High X, Low Y 3700 Y reading Bad
125 Changed Amplifier Output to Slot 2 and messed with DC Voltage
126 P
127 S High X, Low Y 3700 Y reading Bad
128 P GEM 4 <----- PLOTS
129 C 3600 BAD
130 S Center 3700
131 S Low X, Low Y 3700 Y Reading Bad
132 S High X, Mid Y 3700 Y reading good
133 S High X, Mid-Low Y "" Y reading Bad
134 S "" 3750 Y reading Bad
135 S High X, High Y 3700 Y reading good
136 S High X, High Y 3700 Y reading Good, Higher up on Y
162 P GEM 4. Amplification board 1
163 P GEM 4. Amplification board 2, no difference w/ board 2
164 P GEM 4. Amp board 1. changed T.gem.x1.detmap & T.gem.y1.detmap->slot_lo to 1 in database
165 S 3590 bottom->GEM 4 (source). top->GEM2. 2.73mA
166 S 3600 GEM 4
167 S GEM4
168 P cram.gem.crl T1 70->60
169 P cramsTprog T1 60->80
170 P T1 80->70

T

171 P no input connected. Test to see what problem is
172 S Low X, Low Y 3700
173 aborted
174 sS Low X, Low Y 3700 changed to channel 2
175 P replaced Y card and conector
176 P changed CRAM card
177 S Mid X, Mid Y 3700
178 P GEM 2
179 C 3600
180 changed T.gem.X1.detmap & T.gem.Y1.detmap ->slot_lo to 1
181 S changed T.gem.X1.detmap & T.gem.Y1.detmap ->slot_lo to 2 (changed back)
182 S
183 changed T1 to 50 in CRL code for 551b
184 P (kaptons unpluged)
185 P
186 S 3700 Changed T1 from 50 to 40 in CRL code
187 S 3700 CRL code still changed to 40. Data is going to neg(-) input in 550
188 P Changed polarity of trigger. Hold -> no difference
189 S 3700
190 S 3700 Changed CRL T1->70
191 3700 Changed CRL T1->60
192
193 Changed CRL T1-> 50. aborted
194 gate generator to make trigger hold. T1->50
195 S 3700 T1 50 -> 30
196 S 3700 T1 30 -> 20
197 3750 T1 20 -> 70
198 Clear coming from gate generator