Difference between revisions of "Data Log"
From Hall A Wiki
Line 413: | Line 413: | ||
|- | |- | ||
− | | | + | | 199 || P || || || || || |
|- | |- | ||
− | | | + | | 200 || S || || || 3750 || || |
|- | |- | ||
− | | | + | | 201 || || || || || ||looking @ scope |
|- | |- | ||
− | | | + | | 202 || S || || || 3750 || || |
|- | |- | ||
− | | | + | | 203 || || || || || ||looking @ scope |
|- | |- | ||
− | | | + | | 204 || S || || || 3750 || || |
|- | |- |
Revision as of 16:57, 9 September 2010
RUN # | Source/Cosmic/Pedestal | Source Position | Time | HV | # of events | notes |
---|---|---|---|---|---|---|
23 | S | lo x lo y | n/a | n/a | GEM 1 <-----PLOTS | |
24 | S | mid x lo y | n/a | n/a | ||
25 | S | hi x lo y | n/a | n/a | ||
26 | S | hi x mid y | n/a | n/a | ||
27 | S | hi x hi y | n/a | n/a | ||
28 | S | mid x mid y | n/a | n/a | ||
29 | C | n/a | 21 hrs | n/a | 72383 | ROC not responding |
35 | C | n/a | n/a | 3500 | ||
37 | C | n/a | n/a | 3550 | ||
38 | C | n/a | n/a | 3600 | ||
39 | C | n/a | n/a | 3650 | ||
40 | C | n/a | n/a | 3700 | ||
41 | C | n/a | n/a | 3750 | ||
42 | C | n/a | n/a | 3800 | ||
43 | C | n/a | n/a | 3600 | ||
44 | P | n/a | n/a | off | ||
45 | P | n/a | n/a | off | Ped reconfigured | |
46 | S | n/a | n/a | 3600 | 7863 | |
47 | C | n/a | n/a | 3600 | ||
51 | C | n/a | 24.5 hrs | 3600 | 20,000 | |
52 | P | n/a | n/a | off | 11,000 | |
53 | failed | |||||
54 | stopped early | |||||
55 | stopped early | |||||
56 | failed | |||||
57 | 3600 | |||||
58 | C | n/a | 5.2 hrs | 3600 | Client unresponsive | |
59 | S | n/a | n/a | 3600 | ||
60 | S | n/a | n/a | 3600 | ||
74 | S | n/a | n/a | 3700 | ||
75 | P | n/a | n/a | off | GEM # 3 | |
76 | failed | |||||
78 | P | n/a | n/a | off | new read out card | |
80 | C | n/a | n/a | 3700 | ||
81 | S | center | n/a | 3700 | ||
82 | S | center | n/a | 3700 | rate was abnormally high. Reset veto | |
83 | P | n/a | n/a | off | new LV distribution board | |
84 | S | center | n/a | 3700 | ||
85 | Test with VME Crate Transfer - GEM 2 | |||||
86 | "" | |||||
87 | "" | |||||
88 | "" | |||||
89 | "" | |||||
90 | "" | |||||
91 | "" | |||||
92 | "" | |||||
93 | "" | |||||
94 | "" | |||||
95 | "" | |||||
96 | P | |||||
97 | S | No HV - bad | ||||
98 | No Data recorded - Cause was found due to cards with same address | |||||
99 | P | |||||
100 | S | Center | 3700 | GEM 2 <----- PLOTS | ||
101 | S | Low X, Low Y | 3700 | Bad reading in Y | ||
102 | S | High X, High Y | 3700 | No Data recorded - ROC Froze | ||
103 | S | "" | "" | No Data recorded - ROC Froze | ||
104 | S | "" | "" | Reset ROC | ||
105 | S | High X, Low Y | 3700 | Bad reading in Y | ||
106 | S | High X, Mid Y | "" | Y Reading Fine | ||
107 | S | High X, Low-Mid y | "" | Y Reading Bad | ||
108 | P | Changed Y Card - Card was not connected | ||||
109 | P | |||||
110 | S | High X, Low-Mid Y | Y Reading Bad | |||
111 | ABORTED | |||||
112 | S | Center | Old readout Card - It is bad. | |||
113 | S | Center | ROC FROZE | |||
114 | Checked CH1 readout - Nothing | |||||
115 | Checked CH0 | |||||
116 | S | Mid X, High Y | Y reading Fine | |||
117 | P | GEM 3<----- PLOTS | ||||
118 | S | Center | 3700 | |||
119 | S | Low X, Low Y | "" | Y reading bad | ||
120 | S | High X, High Y | "" | Y Reading fine | ||
121 | P | Redid Ped Run | ||||
122 | S | High X, Low Y | 3700 | Y reading Bad | ||
123 | P | Switched Boards on GEM | ||||
124 | S | High X, Low Y | 3700 | Y reading Bad | ||
125 | Changed Amplifier Output to Slot 2 and messed with DC Voltage | |||||
126 | P | |||||
127 | S | High X, Low Y | 3700 | Y reading Bad | ||
128 | P | GEM 4 <----- PLOTS | ||||
129 | C | 3600 | BAD | |||
130 | S | Center | 3700 | |||
131 | S | Low X, Low Y | 3700 | Y Reading Bad | ||
132 | S | High X, Mid Y | 3700 | Y reading good | ||
133 | S | High X, Mid-Low Y | "" | Y reading Bad | ||
134 | S | "" | 3750 | Y reading Bad | ||
135 | S | High X, High Y | 3700 | Y reading good | ||
136 | S | High X, High Y | 3700 | Y reading Good, Higher up on Y | ||
162 | P | GEM 4. Amplification board 1 | ||||
163 | P | GEM 4. Amplification board 2, no difference w/ board 2 | ||||
164 | P | GEM 4. Amp board 1. changed T.gem.x1.detmap & T.gem.y1.detmap->slot_lo to 1 in database | ||||
165 | S | 3590 | bottom->GEM 4 (source). top->GEM2. 2.73mA | |||
166 | S | 3600 | GEM 4 | |||
167 | S | GEM4 | ||||
168 | P | cram.gem.crl T1 70->60 | ||||
169 | P | cramsTprog T1 60->80 | ||||
170 | P | T1 80->70
T | ||||
171 | P | no input connected. Test to see what problem is | ||||
172 | S | Low X, Low Y | 3700 | |||
173 | aborted | |||||
174 | sS | Low X, Low Y | 3700 | changed to channel 2 | ||
175 | P | replaced Y card and conector | ||||
176 | P | changed CRAM card | ||||
177 | S | Mid X, Mid Y | 3700 | |||
178 | P | GEM 2 | ||||
179 | C | 3600 | ||||
180 | changed T.gem.X1.detmap & T.gem.Y1.detmap ->slot_lo to 1 | |||||
181 | S | changed T.gem.X1.detmap & T.gem.Y1.detmap ->slot_lo to 2 (changed back) | ||||
182 | S | |||||
183 | changed T1 to 50 in CRL code for 551b | |||||
184 | P | (kaptons unpluged) | ||||
185 | P | |||||
186 | S | 3700 | Changed T1 from 50 to 40 in CRL code | |||
187 | S | 3700 | CRL code still changed to 40. Data is going to neg(-) input in 550 | |||
188 | P | Changed polarity of trigger. Hold -> no difference | ||||
189 | S | 3700 | ||||
190 | S | 3700 | Changed CRL T1->70 | |||
191 | 3700 | Changed CRL T1->60 | ||||
192 | ||||||
193 | Changed CRL T1-> 50. aborted | |||||
194 | gate generator to make trigger hold. T1->50 | |||||
195 | S | 3700 | T1 50 -> 30 | |||
196 | S | 3700 | T1 30 -> 20 | |||
197 | 3750 | T1 20 -> 70 | ||||
198 | Clear coming from gate generator | |||||
199 | P | |||||
200 | S | 3750 | ||||
201 | looking @ scope | |||||
202 | S | 3750 | ||||
203 | looking @ scope | |||||
204 | S | 3750 | ||||
|