HV and Threshold calibration

From Hall A Wiki
Revision as of 09:30, 13 April 2010 by Rshneor (Talk | contribs)

Jump to: navigation, search

Procedure for the neutron array HV and threshold calibration

Main considerations - 
  • calibrate all neutron bars to equal gains.
  • set gains as low as posibble, to reduce current on PMT's.
  • set threshold energy as low as possible to increase efficiency.
  • set threshold energy high enough to remove low-energy EM gackground.
  • the ADC full range is channels 0-4000.

optimum was chosen at threshold = 5 MeV = 300/4000 channels - need to re-check.

this is equivalent to 1 MeV = 60/4000 channels

The chosen descriminator threshold voltage is 50 mV, which is choose slightly above the hardware limit of 30 mV.


Calibrations steps -

  • Test-area calibration using cosmic muons

Use top-buttom bar trigger, to obtain vertical minimum ionizing muons with known energy deposit.

adjust HV, so that the muon peak will be at-

10.0 cm bar -> E=20 MeV -> channel=1200/4000

12.5 cm bar -> E=25 MeV -> channel=1500/4000

15.0 cm bar -> E=30 MeV -> channel=1800/4000

25.0 cm bar -> E=50 MeV -> channel=3000/4000


  • Beam calibration using the H(e,e'p) reaction.

Use the H(e,e'p) reaction to create 1 GeV/c protons which will have energy loss of 32 MeV for a 10cm wide bar.

Calibrate the gain cofficients (software), so these events will be at channel=1920/4000.

This should be a software calibration, exceptfor bars where the gain correction needed is higher then some factor. In this cases a HV re-calibration will be done first.