HV adjustment
From Hall A Wiki
Revision as of 11:33, 10 November 2010 by Igorko (Talk | contribs) (→High Voltage adjustment software)
High Voltage adjustment software
- During the summer and fall 2010, measurements of gain curves for HAND scintillators were performed in Testlab. Due to the lack of delay cables, electronic units, fastbus crates etc. The measurements were spit in groups of 12 (out of 112).
Almost all measurement were done with same electronics (amplifiers, discriminators...). However, In the Hall we will be using different cables and different electronics. This can cause some disagreement with the measured gain curves.
For this reason following software is prepared. The Algorithm for using this script is following:
1. Calculate the desired voltages for each scintillator based on the Gain Curves. 2. Take a measurement with the calculated voltages. 3. Find the ADC peaks. 4. Run this script with updated information for the included scripts. 5. Apply the new voltages, and check the ADC channel.
6. Make additional iterations if needed