February 12th 2021
From Hall A Wiki
- Updates
- HCAL
- updated to CENTOS 7 , updated library and firmware, one FADC could not be updated by VME, brought to Bill gunning
- issues with decoding fixed by Juan Carlos,
- HCAL
bug in analyzer assumed ROC ID is 16 bit is only 12 bit , needed to mask the sync bit , problem fixed. Sync event was turned on and issue was discovered
- DAQ ready to take cosmics
- VME discriminators ? Hall C has 4 JLab VME discriminators, might be able to borrow 1 or 2
- DAC in HCAL VME crate for other discriminators
- Other option : Brad has a CAMAC crate with discriminators with a few slots available
- could look for already modified NIM discriminators with remote threshold control
- need to run threshold though an insulation module and distribution module - might be in ESB - Brad will check and look.
- Bigbite
- Shower
- Discussion for cabling, baskets that could be craned out for cables. Rack can be unplugged and craned out.
- Shower
HV cables of GRINCH above the BB cables need to coordinate, no need extra basket for HV cables are long and already in basket
- front end BB trigger : Provokar work implementing second trigger, should take a few weeks - Front end Sean, check S7 modules, several broken and are being fixed - need a DAC for Bigbite trigger - Will setup other VME DAQ for BB ( Alex )
- FADC readout ( Mark ) - tried last week, works ok but do not see signal as expected - trigger signal in FADC should see hit for every trigger - window size 500 ns - delay added from TI - ( mode 7 only report one hit - might be better to use mode 1 ) - check using TRig out on SD -
Large offset in preshower channels - need to set thresholds and pedestal in mode 1 to make it works
- Trigger coming from weldment - copy of trigger for WAPP ? Just have the two triggers to TS - Low threshold and high threshold - Low threshold with anticoincidence of high threshold - should be doable in TS directly - need to setup
and test or do logic in NIM and feed to TS
- Hodoscope
- progress on data analysis : several good runs
- Hodoscope
TDC and ADC plots can be seen. Starting to do analysis
- fixing issues seen from data, fixing first and will take more data
- new PHd student helping on cabling diagram
- DAQ issue : no pulse or double pulsing in gate , just need to fix logic
- LHRS
- all crates VME + Fastbus - working now - informations in HALOG - fixes for CODA3 with fastbus - were the fix commited in analyzer ?
- might have been issue with ROC ID ? Will check with Ole at software meeting
- issues with intel CPU, sshd sometimes does not start, reboot with keyboard
- setup portserver ? leave small keyboard and monitor too
- INFN GEM
- problem on 2 MPDs, not producing hits from cosmics - no errors seen and no anomalies in low level test - problem being investigated - chambers were not touched much - might be connection - still to be investigated
- UVA GEM
- couple of bad MPDs, need to have a plan to fix the MPD
- ordering 110 APV cards and 7 MPDs for Moller
- will get 3 with SoLID money
- inventory MPDs done, inventory of bad MPDs
- loan of about 100 APV from Michael Kohl ( some at UVA, all accounted for )
- discussing with GEp to see if we should buy
- TEDF GEM
- working on zero supression - comparing with and without zero suppression
- rate test and latency scans - 5 to 30 us
- working on setting up external clock
- rate tests : no zero supression : 850 Hz ( ethernet data limit )
- rate with zero suppression and 10 % occupancy caps at 2.5 KHz - 50 % occupancy 1 KHZ - still need to understand bottlenecks
- Sean working on understanding cosmics and mapping
- system noisy right now -
- cosmics trigger : first used preshower trigger but not very good signal - adding new scintillator so data would cleaner - might have mapping issues
- clock distribution for MPDs, should do that for the hodoscope - need clock distribution - need to look for one
- CDet
- power supply for NINO - cable length are an issue
- location of LVDS repeaters need to be finalized but needed for GEp5
- VXS crate moved - need to check with William when getting VETOC - have 1 CPU - looking SD
- VETROC : waiting for oscillator - could have 3 available ?
- UVA GEM
- FADC inventory
https://docs.google.com/spreadsheets/d/1DqdMFMznf0gZWxjTXtInLnQpJqaJlRqcUZGdo5cFYwA/edit?usp=sharing
- Hall layout
https://redmine.jlab.org/projects/sbs-daq/wiki/Hall_Pictures