Procedure for the neutron array HV and threshold calibration

Main considerations -

  • calibrate all neutron bars to equal gains.
  • set threshold voltage as low as posibble (50 mV), to reduce current on PMT's.
  • set threshold energy as low as possible to increase efficiency.
  • set threshold energy low enough to be able to separate minimum ionizing electrons.
 (optimum was chosen at around 5 MeV)

We will use the following know energy point for the calibration



Calibrations steps -

  • test-area calibration using cosmic muons

Use top-buttom bar trigger, to obtain vertically going minimum ionizing muons with know energy deposit.

calibrate the muon peak to be at- 10.0 cm bar -> E=20 MeV -> channel=1500 12.5 cm bar -> E=25 MeV -> channel=1875 15.0 cm bar -> E=30 MeV -> channel=2250 25.0 cm bar -> E=50 MeV -> channel=3750