Lab Records

From Hall A Wiki
Revision as of 17:14, 7 December 2010 by Pzhu (Talk | contribs) (New page: === Cell Installed, Arp 04, 2010 === * Name: Boris == Oven Test == == Heater Test == Oven is tested on Mar 24, 2010 without cell. Gas flow is 5~6 in mechanical meter, 508 on digital me...)

(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search

Cell Installed, Arp 04, 2010

  • Name: Boris

Oven Test

Heater Test

Oven is tested on Mar 24, 2010 without cell. Gas flow is 5~6 in mechanical meter, 508 on digital meter.

  • If turn on the constant heater only (set temperature -> 20C), heater temp reached 212C, oven reached 154C

Air Flow meter Calibration @ Mar 9, 2010

Air flow meter in oven control chassis is cross calibrated with mechanical air flow meter at target.

Mechanical Flow Meter Digital Flow Meter
1 260
3 390
5 490
7 600

Alarm 4 is set at 280.