Contents
- 1 Polarization measurement with Hall A Moller polarimeter
- 2 Check list before measurements
- 2.1 Helicity, BCM signals in Hall A Counting Room
- 2.2 Helicity, BCM, 100kHz signals in Hall A Moller electronics rack
- 2.3 Helmholtz coils power supply
- 2.4 Target motion
- 2.5 Start CODA Moller database (msqld)
- 2.6 CODA LED test run
- 2.7 Check Helicity signals and scalers data in CODA data file
- 2.8 ADC Pedestals check
- 3 Software to start
- 4 Moller magnet settings and beam tuning
- 5 Detector HV tuning
- 6 Thresholds tuning
- 7 Polarization measurements
- 8 Results presentation
- 9 Spin dance
- 10 Troubleshooting
Polarization measurement with Hall A Moller polarimeter
Check list before measurements
Helicity, BCM signals in Hall A Counting Room
Patch panel in Counting Room with cables going to the Hall A:
Connector# Signals (all NIM level) (width) 4 - BCM signal for Moller (1us) from VtoF unit (11) Dx1 14 - Helicity signal (33ms) ----_____--- 15 - QRT signal (33ms) ----_____---- 16 - MPS signal (200 us) ---__---- front with 1us ahead of front of Helicity signal
Helicity, BCM, 100kHz signals in Hall A Moller electronics rack
- NIM Module type of PS757 (Fan in/out) in NIM crate of FADC electronics rack
#input Signals width 1 Clock 100kHz (50ns) 4 Helicity (33ms) 7 MPS (200us)
- ADC V792 in FADC VME crate:
#CH 00 Helicity signal #CH 01 QRT signal GATE 1.5us after front of MPS signal, width=130ns
- Check APERTURE THRESHOLD with digital multimeter on NIM 16ch discriminator module type of PS708 in NIM crate before old DAQ rack.
APERTURE THRESHOLD = -300mV (readout with digital multimeter)
- Check calorimeter OFFSETs with digital multimeter on NIM modules type of PS748 in slot #1 and PS740 slot #2 NIM crate in old DAQ rack.
- Nov 25, 2014
PS748 (slot #1) Ch # OFFSET(mV) 0 +18.5 1 +18.6 2 +20.4 3 +21.7 4 +19.5 5 +25.5 6 +19.3 7 +10.3
PS740 (slot #2) Ch # OFFSET(mV) 0 +21.5 1 +21.6 2 +22. 3 +21.5
Helmholtz coils power supply
Target motion
Start CODA Moller database (msqld)
Login to adaql1 as moller and check if msqld daemon is running:
adaql1> ps -u moller | grep msqld
if output like that:
22600 ? 00:00:01 msqld
that means msqld is running.
If not, to start it type:
adaql1> run_daemon_db msqld &
CODA LED test run
Check Helicity signals and scalers data in CODA data file
To check helicity, QRT signals type in paw:
paw> exe run run=15623 Run=15624 is attached to lun=21 paw> nt/sca //lun21/1 nsca>0 50 1 ! itick itrig(1) itrig(6) itrig(8)
This is output for 30Hz helicity (pair):
+-------+-------------+-------------+-------------+-------------+ | Event | itick | itrig(1) | itrig(6) | itrig(8) | +-------+-------------+-------------+-------------+-------------+ | 2 | 11155369 | z'0' | z'1' | z'0' | | 5 | 11155373 | z'0' | z'0' | z'1' | | 8 | 11155377 | z'0' | z'0' | z'1' | | 11 | 11155381 | z'0' | z'1' | z'1' | | 14 | 11155385 | z'0' | z'0' | z'1' | | 17 | 11155389 | z'0' | z'1' | z'1' | | 20 | 11155393 | z'0' | z'1' | z'1' | | 23 | 11155397 | z'0' | z'0' | z'1' | | 26 | 11155401 | z'0' | z'0' | z'0' | | 29 | 11155405 | z'0' | z'1' | z'1' | | 32 | 11155409 | z'0' | z'1' | z'1' | | 35 | 11155413 | z'0' | z'0' | z'1' | | 38 | 11155417 | z'0' | z'1' | z'1' | | 41 | 11155421 | z'0' | z'0' | z'1' | | 44 | 11155425 | z'0' | z'0' | z'1' | | 47 | 11155429 | z'0' | z'1' | z'1' | | 50 | 11155433 | z'0' | z'1' | z'0' | +-------+-------------+-------------+-------------+-------------+
Where:
itick - VxWorks time ticks (set in ~/daq/coda2/crlppc/moller_onl.crl to 8.33mS) itrig(1) - ADC gate on input #1 of TIR board itrig(6) - Helicity signal on input #6 of TIR board itrig(8) - QRT signal on input #8 of TIR board
To check scalers data type in paw:
paw> nt/sca //lun21/1 nsca>0 50 1 ! itick isca(1) isca(2) isca(3) isca(4) isca(5) +-------+-------------+-------------+-------------+-------------+-------------+-------------+ | Event | itick | isca(1) | isca(2) | isca(3) | isca(4) | isca(5) | +-------+-------------+-------------+-------------+-------------+-------------+-------------+ | 2 | 11155369 | 9315 | 9315 | 9315 | 0 | 0 | | 5 | 11155373 | 9463 | 9463 | 9463 | 0 | 0 | | 8 | 11155377 | 9611 | 9611 | 9611 | 0 | 0 | | 11 | 11155381 | 9760 | 9760 | 9760 | 0 | 0 | | 14 | 11155385 | 9909 | 9909 | 9909 | 0 | 0 | | 17 | 11155389 | 10057 | 10057 | 10057 | 0 | 0 | | 20 | 11155393 | 10205 | 10205 | 10205 | 0 | 0 | | 23 | 11155397 | 10353 | 10353 | 10353 | 0 | 0 | | 26 | 11155401 | 10502 | 10502 | 10502 | 0 | 0 | | 29 | 11155405 | 10650 | 10650 | 10650 | 0 | 0 | | 32 | 11155409 | 10798 | 10798 | 10798 | 0 | 0 | | 35 | 11155413 | 10946 | 10946 | 10946 | 0 | 0 | | 38 | 11155417 | 11095 | 11095 | 11095 | 0 | 0 | | 41 | 11155421 | 11243 | 11243 | 11243 | 0 | 0 | | 44 | 11155425 | 11391 | 11391 | 11391 | 0 | 0 | | 47 | 11155429 | 11539 | 11539 | 11539 | 0 | 0 | | 50 | 11155433 | 11688 | 11688 | 11688 | 0 | 0 | +-------+-------------+-------------+-------------+-------------+-------------+-------------+
Where:
isca(1) - Left arm (LG and Ap) isca(2) - Right arm (LG and Ap) isca(3) - Coincidence Left and Right arms isca(4) - Accidentals Left and Right arms isca(5) - BCM isca(6) - PLU out 7; in LED setup - number of LED flashes isca(7) - isca(8) - isca(9) - isca(10) - (Helicity delayed)*Clock(100kHz) isca(11) - !MPS isca(12) - PLU out #6 (TimerGenarator * helicity window) isca(13) - isca(14) - isca(15) - Clock TimerGenerator 100kHz
ADC Pedestals check
- Set HV ON on moller detector. (Need link to HV ON)
- Select configuration "moller_puls" in old CODA, download it.
- In "mpc" programm load configuration "PULSER_SET" to detector electronics and start CODA run.
- Take 3000-5000 events and stop run.
- In ~/paw/moller start paw:
> paw paw> exec run run=15627 paw> exec lg_pedes run=15627 pulschk=1
- Copy output results into text file with name "pedes_#RunNumber.dat", where #RunNumber is CODA "pulser" run.
- Edit file "lg_spectra.kumac", replace:
... else ve/read pedes,pedcut pedes_14754.dat endif ...
with:
elseif [run]<#RunNumber then ve/read pedes,pedcut pedes_14754.dat else ve/read pedes,pedcut pedes_#RunNumber.dat endif ...
Software to start
- CODA
- old DAQ ( data in /adaq1/data1/raw ):
ssh moller@adaql1 adaql1> start_coda
- Detector settings and HV control GUI
ssh moller@adaql2 adaql2> cd Java/msettings/ adaql2> mpc
- FADC DAQ ( data in /data1/raw ) :
ssh moller@hamoller (open 3 terminals) hamoller> start_coda_brad
- FADC settings:
hamoller> cd crl/FADC/ hamoller> gui_fadc_par
- FADC online monitor:
hamoller> cd fadc_online/mollmon/ hamoller> mollmon -n moller25 -h hamoller
- EPICS control
ssh adaq@adaql1 (or adaq1) adaql1> monticello (OS Menu -> EDM -> Hall A -> HallA Moller polarimeter control)
OR
adaql1> jtabs (Beam Setup -> HallA-> Magnets-> Moller-> Combo)
Moller magnet settings and beam tuning
- Ask target operator to remove the main target
- Raster OFF
- Moller magnet settings: (Q1,Q2,Q3,Q4, Dipole-OFF)
1st Moller quadrupole magnet Q1 - MQO1H02 2nd Moller quadrupole magnet Q2 - MQM1H02 3rd Moller quadrupole magnet Q3 - MQO1H03 4th Moller quadrupole magnet Q4 - MQO1H03A Moller dipole magnet - MMA1H01
- Get magnet settings for energy (as example) e0=7.37GeV
paw> lcd mag_optim paw> exec sett_magp12 e0=7.37 Eb=7.37GeV . Q1 Q2 Q3 Q4 Dipole kG -4.2742 -1.85808 1.994 4.345 10.8837 kG*m -30791.3 -16358.5 14315.3 30839.5 1.78793E+06 A -197.853 -93.5748 97.3212 208.451 399.068
- Ask MCC operator for beam current 0.5uA
- Beam position tuning with magnet settings (Q1,Q2,Q3,Q4) on IPM1H01 BPM ~0.2mm (absolute X,Y)
- Turn Beam OFF
- Turn ON Moller Dipole
- Ask Beam current 0.5uA
- Beam position tuning with Dipole ON on IPM1H01 BPM ~0.2mm
- Set target on beam
1: SM 13.0um (old), 2: SM 29.4um, 3: Fe 14.3um, 4: Fe 9.3um, 5: SM 6.8um, Left.Lim: Al 16.6um
Detector HV tuning
- HV for Moller detector:
portserver: hatsv5:2003 slot #5: ch:(0-3) - Calorimeter Left (1-4); ch:(4-7) - Calorimeter Right (1-4) slot #6: ch:(0-3) - Aperture Left (1-4); ch:(4-7) - Aperture Right (1-4)
- HV prediction (printout predicted HV calorimeter values for energy e0):
paw> exec hvlg e0=7.3 set=0
- Tune detector HV
- Start CODA run
- Check signals peak position at ADC channel.
paw> exec run run=15625 paw> exec lg_spectra run=15625
- Change HV for calorimeter in order to align the Møller peak position at ADC channel 300 for each block.
- Repeat from #1
- FADC DAQ: check signals peak position on online monitor 'mollmon'
Thresholds tuning
Polarization measurements
- Start dbedit on adaql1, select "mollerpol_l" database and set in "beam_pol" configuration table in row "code" parameter "adcevents" from 9 to 2: " ... adcevents=2"
- Make at least two CODA runs in order to use both coil polarities.
- Data analysis:
- old DAQ: adaql1:~/paw/moller/
> paw paw> exec run run=15625 paw> exec lg_spectra run=15625 paw> exec time run=15625 paw> exec asym11 run=15625
- FADC DAQ: hamoller:~/fadc_analysis/moller_fadc-15Nov2011/
> bash > source root-setup.sh > cd onlana > analyser analyser> replay_test(3023) analyser> .x do_analyse(3023)
- old DAQ: adaql1:~/paw/moller/
Results presentation
- Results of all runs from "asym11" stored in files: runs.tab, res.dat
- Do copy these files to paw/moller/meas:
> cp runs.tab meas/runs.tab.#RunBeg-#RunEnd > cp res.dat meas/res.dat.#RunBeg-#RunEnd
- To plot and average results, edit file: ~/paw/moller/asym_avt.kumac
- comment old beam polarization data and add new one, modify output of 'ps' file and titles with date then run it.
Spin dance
Troubleshooting
- VME crates reboot
- Software reboot if VME crate is running:
- old DAQ hallavme5:
> telnet hatcv5 2004 > reboot Ctrl and "]" telnet> quit
- FADC DAQ:
> telnet hatsv12 2005 > rebootrebooting VME crate Ctrl and "]" telnet> quit
- old DAQ hallavme5:
- Hardware reboot if crates not response:
- On adaql2 start firefox
old DAQ hallavme5:
Go to: http://hareboot6
(outlet #8)user: hlauser select "Device Manager" tab, select "Control" from left menu, select in small box Outlet #8, select Action "Reboot Immediatly " and press "Next" button to activate action.
Wait for few minutes while VME crate is rebooting.
- FADC DAQ crate:
Go to: http://hareboot26
(outlet #1)user: hlauser select in "ON" box Outlet #1, select in small box Outlet #1, select Action "Reboot Immediatly " and press "Next" button to activate menu. press "Apply" button to activate action.
Wait for few minutes while VME crate is rebooting.
- On adaql2 start firefox
old DAQ hallavme5:
- HV control